Electron beam monitoring of a modified conventional medical accelerator with a portable current transformer system traceable to primary electrical standards.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Characterization of a shielded beam current transformer for ultra-high dose rate (FLASH) electron beam monitoring and dose reporting.Medical physics · 2025Article
- Electron beam monitoring of a modified conventional medical accelerator with a portable current transformer system traceable to primary electrical standards.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundUltra-high dose rate radiotherapy (UHDR) delivers therapeutic doses at rates >40 Gy/s in a fraction of a second, aiming to enhance the therapeutic ratio through the FLASH effect. The substantial increase in UHDR beam current poses serious challenges for conventional active dosimeters. Integrating current transformers (ICT) offer a nondestructive solution for accurate monitoring, enabling the type of fast transient readout that will be crucial for UHDR treatment verification. PURPOSE: The aim of this study is to build and characterize a clinically deployable ICT system and to develop an accurate calibration methodology for absolute charge determination that is traceable to primary electrical standards.
methodsThe ICT was constructed from a Super MuMetal
resultsThe ICT responses for nominal test pulses, generated by a function generator and a current pulser, exhibited accurate reproduction of rise and fall times within the 1 ns sampling frequency. A systematic droop effect of 0.6(1)%/µs was observed but is accounted for through the calibration chain. The calibration of the current pulser exhibited a repeatability typically better than 0.05%, with a slowly-varying leakage that can be subtracted using a linear regression of the leakage current. The ICT charge calibration demonstrated a repeatability in the range of (0.5 to < 0.05)% for charge per pulse values in the range of (0.5 to 50) nC, respectively. The ICT response showed a strong linear relationship (adj-R
conclusionsThis work demonstrates the feasibility of accurately calibrating the ICT in terms of absolute charge and applying it as a clinically deployable monitoring system of mA-scale electron beams delivered by a medical accelerator.