Solid-state esophageal pressure sensor for the estimation of pleural pressure: a bench and first-in-human validation study.
Article in Critical care (London, England), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05817968 (Comparison of a Solid State Versus Balloon Esophageal Catheter for Estimation of Pleural Pressure in Surgical ICU Patients), which is not on this map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Comparison of a Solid State Versus Balloon Esophageal Catheter for Estimation of Pleural Pressure in Surgical ICU Patients
Who cites it
1 citing paper in PubMed.
- Assessing inspiratory drive and effort in critically ill patients at the bedside.Critical care (London, England) · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAdvanced respiratory monitoring through the measurement of esophageal pressure (Pes) as a surrogate of pleural pressure helps guiding mechanical ventilation in ICU patients. Pes measurement with an esophageal balloon catheter, the current clinical reference standard, needs complex calibrations and a multitude of factors influence its reliability. Solid-state pressure sensors might be able to overcome these limitations.
objectivesTo evaluate the accuracy of a new solid-state Pes transducer (Pes
methodsAbsolute and relative solid-state sensor Pes measurements were compared to a reference pressure in a 5-day bench setup, and to simultaneously placed balloon catheters in 15 spontaneously breathing healthy volunteers and in 16 mechanically ventilated ICU patients. Bland-Altman analysis was performed using mixed effects modelling with bootstrapping to estimate bias and upper and lower limits of agreement (LoA) and their confidence intervals.
resultsBench study: Solid-state pressure transducers had a positive bias (P
conclusionsThe novel solid-state pressure transducer showed good accuracy on the bench, in healthy volunteers and in ventilated ICU-patients. This could contribute to the implementation of Pes as advanced respiratory monitoring technique.
trial registrationClinicaltrials.gov identifier: NCT05817968 (patient study). Registered on 18 April 2023.