Evidence map›Paper›PMID 39807094›Full record

ArticleJournal of experimental orthopaedics2025

Navigated instrumentation and ligament tensioning device enhances initial gap acquisition during total knee arthroplasty procedure: A cadaveric study.

François Boux de Casson, Laurent Angibaud, Florian Kerveillant, Faustine Nogaret, Joris Ruffin, Léonard Duporté, Gérard Giordano, Louis Dagneaux

Abstract read
In one paragraph

Article in Journal of experimental orthopaedics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

François Boux de CassonBlue-Ortho Meylan France.ORCID 0000-0001-7567-2651
Laurent AngibaudExactech Inc Gainesville Florida USA.ORCID 0009-0006-5291-0155
Florian KerveillantBlue-Ortho Meylan France.
Faustine NogaretDepartment of Orthopaedic Surgery and Trauma University Center of Montpellier, University of Montpellier Montpellier France.
Joris RuffinDepartment of Orthopaedic Surgery and Trauma University Center of Montpellier, University of Montpellier Montpellier France.
Léonard DuportéDepartment of Orthopaedic Surgery and Trauma University Center of Montpellier, University of Montpellier Montpellier France.ORCID 0000-0002-2691-7197
Gérard GiordanoJoseph Ducuing Hospital Toulouse France.ORCID 0000-0002-9180-8448
Louis DagneauxDepartment of Orthopaedic Surgery and Trauma University Center of Montpellier, University of Montpellier Montpellier France.ORCID 0000-0001-7629-5560

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: Gap-balanced total knee arthroplasty (TKA) technique relies on initial ligament evaluation, particularly in patient-specific implantation using computer-assisted technologies. This cadaveric study aimed to compare the reproducibility and reliability of medial and lateral gap measurements between manual stress testing and dynamic ligament balancer. Methods: Initial gap acquisitions were assessed from eight cadaveric knees (four specimens) during the same navigated TKA procedure by five differently skilled surgeons (three seniors and two juniors). Medial and lateral gaps were sequentially acquired from extension to maximum knee flexion, applying manual stress prior to any bone cuts (conventional technique), and using intra-articular tensioning device placed between the tibial cut and the native femur (instrumented technique). Reproducibility was assessed using intraclass correlation coefficient (ICC), stratified by the measurement technique, the type of gaps and the operator experience. Differences in gaps (mm) between techniques were assessed using the Bland and Altmann method. Results: The instrumented technique showed higher ICCs than the conventional technique for medial and lateral gaps (0.87 vs. 0.60, Conclusion: The instrumented balancing technique offered better reproducibility than using manual valgus and varus stress, when measuring medial and lateral gaps. Tensioning devices may play a significant role in enhancing initial gap acquisition, disregarding the flexion angle and the operator experience. Level of Evidence: Level IV (observational study involving cadaveric specimens).

Indexed as

alignmentexperimental modeljoint laxitykinematicsligamenttensorTKA

Identifiers

PMID39807094
PMCPMC11725761

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.