ArticleSensors (Basel, Switzerland)2025

Fiber Bragg Grating Thermometry and Post-Treatment Ablation Size Analysis of Radiofrequency Thermal Ablation on Ex Vivo Liver, Kidney and Lung.

Sanzhar Korganbayev, Leonardo Bianchi, Clara Girgi, Elva Vergantino, Domiziana Santucci, Eliodoro Faiella, Paola Saccomandi

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Sanzhar KorganbayevDepartment of Mechanical Engineering, Politecnico di Milano, Via Giuseppe La Masa 1, 20156 Milan, Italy.ORCID 0000-0002-6299-6048
Leonardo BianchiDepartment of Mechanical Engineering, Politecnico di Milano, Via Giuseppe La Masa 1, 20156 Milan, Italy.ORCID 0000-0001-6100-0862
Clara GirgiDepartment of Mechanical Engineering, Politecnico di Milano, Via Giuseppe La Masa 1, 20156 Milan, Italy.
Elva VergantinoOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo 200, 00128 Rome, Italy.ORCID 0009-0008-6968-0645
Domiziana SantucciOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo 200, 00128 Rome, Italy.ORCID 0000-0003-4903-8715
Eliodoro FaiellaOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo 200, 00128 Rome, Italy.ORCID 0000-0002-7367-6891
Paola SaccomandiDepartment of Mechanical Engineering, Politecnico di Milano, Via Giuseppe La Masa 1, 20156 Milan, Italy.ORCID 0000-0003-4236-8033

Funding

European Research Council 759159
6 · The paper itself

Abstract

Radiofrequency ablation (RFA) is a minimally invasive procedure that utilizes localized heat to treat tumors by inducing localized tissue thermal damage. The present study aimed to evaluate the temperature evolution and spatial distribution, ablation size, and reproducibility of ablation zones in ex vivo liver, kidney, and lung using a commercial device, i.e., Dophi™ R150E RFA system (Surgnova, Beijing, China), and to compare the results with the manufacturer's specifications. Optical fibers embedding arrays of fiber Bragg grating (FBG) sensors, characterized by 0.1 °C accuracy and 1.2 mm spatial resolution, were employed for thermometry during the procedures. Experiments were conducted for all the organs in two different configurations: single-electrode (200 W for 12 min) and double-electrode (200 W for 9 min). Results demonstrated consistent and reproducible ablation zones across all organ types, with variations in temperature distribution and ablation size influenced by tissue characteristics and RFA settings. Higher temperatures were achieved in the liver; conversely, the lung exhibited the smallest ablation zone and the lowest maximum temperatures. The study found that using two electrodes for 9 min produced larger, more rounded ablation areas compared to a single electrode for 12 min. Our findings support the efficacy of the RFA system and highlight the need for tailored RFA parameters based on organ type and tumor properties. This research provides insights into the characterization of RFA systems for optimizing RFA techniques and underscores the importance of accurate thermometry and precise procedural planning to enhance clinical outcomes.

Indexed as

KidneyLiverLungRadiofrequency AblationThermometryAnimalsOptical Fibersfiber Bragg grating sensorskidneyliverlungradiofrequency ablationthermometry