ArticleJournal of applied clinical medical physics2025
Validation of a diode-based phantom for high temporal and spatial measurements in a 1.5 T MR-linac.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Real-time dose reconstruction and dose coverage forecasting using the magnetic resonance linear accelerator.Physics and imaging in radiation oncology · 2026Article
- Validation of a diode-based phantom for high temporal and spatial measurements in a 1.5 T MR-linac.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundFor the development and validation of dynamic treatment modalities and processes on the MR-linac, independent measurements should be performed that validate dose delivery and linac behavior at a high temporal resolution. To achieve this, a detector with both high temporal and spatial resolution is necessary. PURPOSE: This study investigates the suitability of a Delta4 Phantom+ MR (Delta4) detector array for time-resolved dosimetry in the 1.5 T MR-linac and characterizes the Delta4's performance with dynamic dose delivery, such as beam gating and field modulation during radiation.
methodsA Delta4 detector was used, including software for time-resolved dosimetry. First validation experiments were performed and compared to reference measurements. Subsequently, demonstrator measurements were performed to show use cases of the Delta4's time-resolved dose readouts. An example of such an experiment is the determination of the field speed during a sliding window experiment, traveling between 0.7 and 6.8 cm/s in the cranial-caudal direction.
resultsValidation experiments of the dose reproducibility and dose rate dependency showed no difference relative to the standard static delivery. The field speed measured by the Delta4 showed an average field speed difference of -0.3% relative to MR-linac log files. The Delta4 was capable of measuring the dose with high accuracy and temporal resolution during dynamic radiation delivery.
conclusionThe Delta4 can be used for time-resolved dosimetry in a 1.5 T MR-linac.