ArticleThe international journal of medical robotics + computer assisted surgery : MRCAS2024
Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case-Series Study.
Article in The international journal of medical robotics + computer assisted surgery : MRCAS, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Limited Accuracy of Tibial Stylus Measurements in Conventional Total Knee Arthroplasty.Journal of clinical medicine · 2026Article
- Accuracy and Precision of Femoral and Tibial Bone Resections Using Manual Unrestricted Kinematic Alignment in Total Knee Arthroplasty: A Retrospective Digital Caliper Study.Journal of clinical medicine · 2026Article
- A Novel Mixed-Reality System Has Comparable Accuracy to a Robotic System in Total Knee Arthroplasty.Arthroplasty today · 2026Article
- External femoral rotation of 2 degrees is associated with the lowest stuffing rates around the femur in functionally aligned TKA.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2025Article
- Femoral resection accuracy and precision in manual caliper-verified kinematic alignment total knee arthroplasty.Journal of experimental orthopaedics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundWe investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system.
methodsCalliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024.
resultsPlanned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) (p = 0.217) and 0.94 mm ± 1.15 mm (p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm (p = 0.001) and 0.89 mm ± 0.8 mm (p = 0.008) for medial and lateral PT, and 1.1 mm ± 0.97 mm (p = 0.001) and 1.04 mm ± 0.79 mm (p = 0.001) for medial and lateral PC, respectively.
conclusionThe ROSA robotic system can achieve a high degree of accuracy for planned resection thickness. Results are valid only for the imageless ROSA TKA in patients with primary knee osteoarthritis.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials