Commissioning of the Varian universal interstitial cylinder system for HDR brachytherapy of gynecological cancer.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Commissioning of the Varian universal interstitial cylinder system for HDR brachytherapy of gynecological cancer.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis paper outlines the commissioning of the Varian (VMS, Varian Medical Systems, Palo Alto, CA) Universal Interstitial Cylinder (UIC) applicator set for Ir-192 HDR brachytherapy. The UIC was commissioned for use with CT and MRI and a custom phantom was designed to avoid the introduction of water-like materials into the needle guide tracks. Various marker strands were investigated to determine which allowed the most accurate reconstruction of source positions.
methodsPlanar kV and MV imaging, along with physical measurements and autoradiographs, were used to commission the physical dimensions of all components of the UIC applicator set. CT and MR imaging was used to further commission one configuration of the UIC with UCP and eight interstitial needles in a simulated clinical setup using a GYN phantom. Three different methods of channel identification were compared - no radio opaque markers, VMS numbered markers, or nylon coated stainless steel leader wires - to see which best aided in channel identification and image registration. An HDR MRI Lumen marker (C4 Imaging, LLC) was used to verify any applicator rotation on MR scans during image registration. Three types of GYN phantoms were investigated - wet towel, gelatin, and ground beef. Dimensions of all components were compared with vendor provided information, including the solid applicator models, which are based on the computer-aided design model files of the specific applicators.
resultsThe dimensions of the applicators could be validated using physical measurements, kV and MV planar imaging, and CT scans. The ground beef based GYN phantom best eliminated the introduction of water into the needle guide tracks that was found when using a water or gel-based phantom. CT scans using no radio opaque markers did not allow the plastic needles to be visualized well enough to digitize source positions. CT scans with VMS markers showed significant artifact. CT scans with the nylon coated stainless steel wires provided the best visibility of the needle locations to aid in digitizing source positions. The use of an MR marker allowed the channel to be identified on the MR scan and confirm rotation for image registration.
conclusionsThe UIC set and applicator configuration was commissioned for CT and MR based treatment planning. The plastic components of the UIC applicator set pose challenges to the commissioning process but the use of radio opaque markers seen on CT combined with MR image registration allow the source positions within the needles, as well as the location of the end of the needles, to be digitized appropriately. A ground beef phantom minimized the fluid introduced into the needle guide track, minimizing any unintended MR and CT signal in the needle guide tracks.