The Development of a Forceps-Adaptable Pressure Device for Instrumental Delivery: A Proof-of-Concept Study for Clinical and Educational Applications.
Article in Sensors (Basel, Switzerland), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo develop and validate a device that measures the pressure exerted by forceps on the fetal head for clinical use.
backgroundThe lack of clinical tools to quantify forceps pressure on the fetal head may impact maternal and neonatal outcomes. Existing studies have not measured the direct contact pressure between forceps blades and the fetal head, highlighting the need for innovation.
methodsWe integrated fluid pressure transducers into obstetric forceps using fluid-filled tubing encased in flexible silicone socks attached to the blades. Tubing materials-polyvinyl chloride (PVC) and polyurethane (PU)-and fluids (air and water) were tested with both biocompatible and non-biocompatible silicone socks. An onboard electronic board collected pressure data and transmitted them via Bluetooth for real-time analysis. The system was evaluated on a custom-built bench simulating forceps application.
resultsAir-filled tubing exhibited significant drift and low accuracy due to air compressibility. Water-filled PU tubing reduced drift but was still suboptimal. Water-filled PVC tubing with both types of silicone socks provided the best results, showing minimal drift and a strong correlation between measured pressures and applied forces.
conclusionsThe developed device represents a significant advancement, as no existing system measures the pressure exerted by forceps blades on the fetal head. By effectively measuring pressure across the entire contact surface in real time, it offers applications in both training and clinical practice. The device allows for objective feedback, potentially improving the safety and efficacy of forceps deliveries. Future work includes comprehensive mannequin tests and eventual in vivo studies to validate its effectiveness in realistic settings, aiming to enhance obstetric training and reduce maternal and neonatal complications.