Evidence map›Paper›PMID 39666089›Full record

ArticleArchives of orthopaedic and trauma surgery2024

Accuracy of leg length changes in total hip arthroplasty using a computed tomography-based augmented reality navigation system.

Gai Kobayashi, Shintaro Ichikawa, Shine Tone, Yohei Naito, Akihiro Sudo, Masahiro Hasegawa

Abstract read
PubMed Publisher
In one paragraph

Article in Archives of orthopaedic and trauma surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Gai KobayashiDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Shintaro IchikawaDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Shine ToneDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Yohei NaitoDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Akihiro SudoDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Masahiro HasegawaDepartment of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan. masahase@clin.medic.mie-u.ac.jp.ORCID http://orcid.org/0000-0001-7948-2003

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionLeg length is an important consideration in total hip arthroplasty (THA) as leg length discrepancies (LLD) after THA cause poor outcomes and medical litigation. This study aimed to investigate the accuracy of computed tomography (CT)-based navigation with augmented reality (AR) technology for measuring intra-operative leg length change using anteroposterior radiography (two-dimensional, 2D) and CT (three-dimensional, 3D). MATERIALS AND

methodsThis study included 71 patients (75 hips) who underwent primary cementless THA in the supine position between June 2022 and November 2023 using the HoloNavi One. To assess the accuracy of the HoloNavi One based on 2D and 3D measurements, the absolute error between the intra-operative leg length change and the 2D and 3D measurements was evaluated, and the correlations were investigated. Additionally, factors affecting the absolute error were examined using multivariate analysis.

resultsThe mean intra-operative leg length change when using the HoloNavi One was 6.5 ± 5.0 mm, while the mean leg length change on the 2D and 3D measurements were 5.9 ± 4.8 mm and 4.6 ± 5.7 mm, respectively. The mean absolute errors were 2.3 ± 2.7 mm between the HoloNavi One and 2D measurements, and 3.8 ± 3.3 mm between HoloNavi One and 3D measurements. The absolute errors in leg length changes for the 3D measurements were greater than those for the 2D measurements (p < 0.01). Positive correlations of leg length changes were found between the HoloNavi One and the 2D and 3D measurements. In the multiple regression analysis, no significant factors affecting the absolute error were identified in either the 2D or 3D measurements.

conclusionsCT-based navigation with AR technology in the supine position provided acceptable accuracy for leg length change measurements.

Indexed as

Arthroplasty, Replacement, HipAugmented RealityLeg Length InequalityTomography, X-Ray ComputedAdultAgedAged, 80 and overFemaleHumansImaging, Three-DimensionalMaleMiddle AgedRetrospective StudiesSurgery, Computer-AssistedSurgical Navigation SystemsAugmented reality technologyComputed tomography-based navigationLeg lengthThree-dimensionalTotal hip arthroplastyTwo-dimensional