Evidence map›Paper›PMID 39664255›Full record

ArticleOrthopaedic journal of sports medicine2024

Optimizing Reduction Guide Stability in Osteotomy Using Patient-Specific Instrumentation: A Basic Guideline.

Michel Meisterhans, Christoph Zindel, Bastian Sigrist, Sandro F Fucentese, Lazaros Vlachopoulos

Abstract read
In one paragraph

Article in Orthopaedic journal of sports medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Michel MeisterhansDepartment of Orthopaedics, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
Christoph ZindelDepartment of Orthopaedics, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.ORCID https://orcid.org/0000-0002-6253-9824
Bastian SigristResearch in Orthopedic Computer Science (ROCS), Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
Sandro F FucenteseDepartment of Orthopaedics, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
Lazaros VlachopoulosDepartment of Orthopaedics, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: The use of patient-specific instruments (PSIs) for osteotomies is becoming more popular in orthopaedic surgery for correcting mechanical axis and posttraumatic deformities. However, the PSI reduction guides have great potential for intraoperative deformation, which adversely affects the accuracy of the procedure. Purpose: To conduct a finite element analysis (FEA) to analyze different design parameters to improve the intraoperative stability of the reduction guides. Study Design: Descriptive laboratory study. Methods: A reduction guide with a rectangular cross section and four 4-mm K-wire slots was simplified, and the following parameters were modified: width, height, profile design, K-wire thickness, and positions. Bending and torsional moments were applied to the guide construct and guide deformation and equivalent stress were determined using FEA. Results: Increasing the profile height by 25% resulted in a 44% reduction in guide deformation for bending (37% for torsion). A 25% increase in profile width led to an 18% deformation reduction for bending (22% for torsion). Transverse K-wire slots resulted in 51% less deformation in torsion compared with longitudinally oriented slots. Placing the central K-wire slots 25% closer to the osteotomy reduced guide deformation by 20% for bending and 11% for torsion. Conclusion: The most effective methods to increase reduction guide stability are to increase the guide height and reduce the central K-wire distance to the osteotomy. Clinical Relevance: When performing opening or closing wedge osteotomies, which mainly involve bending of the guide, a high-profile guide and longitudinally oriented K-wire slots should be used. When torque is expected as in rotational osteotomies, the K-wire holes in guides should be oriented transversely to reduce intraoperative deformation.

Indexed as

biomechanicsbiomechanics of bonefinite element analysiskneeosteotomyPSIreduction guidestability