Evidence map›Paper›PMID 39640317›Full record

ArticleSurgical neurology international2024

Biological and hardware-related spinal cord stimulation complications and their management: A single-center retrospective analysis of the implantation of nonrechargeable implantable pulse generators in different pain conditions.

Marek Prokopienko, Michał Sobstyl

Abstract read
In one paragraph

Article in Surgical neurology international, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Marek ProkopienkoDepartment of Neurosurgery, Institute of Psychiatry and Neurology, Warsaw, Poland.
Michał SobstylDepartment of Neurosurgery, Institute of Psychiatry and Neurology, Warsaw, Poland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: We present our experience with spinal cord stimulation (SCS) for patients suffering from different pain conditions who subsequently developed hardware-related complications after SCS surgery. The SCS hardware-related complications may compromise the continuous SCS therapy due to partial or total hardware removal. Such situations should be avoided, and possible predisposing factors for their development should be minimized. The present study aimed to evaluate the frequency of hardware-related complications and their proper neurosurgical management. Methods: The study is designed as a retrospective analysis of all hardware-related complications of SCS procedures for pain patients who underwent the implantation of the nonrechargeable PrimeAdvanced Results: Among 20 patients, 8 (40%) patients were affected by hardware-related complications. The most common complications were skin erosion found in 5 patients (25%) and incorrect functioning of the implantable pulse generator (IPG) affecting 2 patients (10%). There were 1 case of an IPG migration (5%) and 1 hardware infection (5%) due to a staphylococcal wound. A total number of 16 revision surgeries were performed to manage all hardware-related complications in these patients adequately. Most of the patients (5 of them) were troubled by more than one hardware-related complication episode. Three patients had 3 revision surgeries, 2 patients had 2 revision surgeries, and 3 patients had 1 revision surgery. Among 8 patients with complications, 3 patients had no further continuation of SCS therapy due to hardware-related complications. Among these 3 patients who stopped their SCS therapy, 1 patient had 3 hardware-related episodes, and the remaining 2 patients were troubled by two hardware-related episodes before discontinuation of SCS therapy. Conclusion: Our results indicate that patients treated by the SCS technique are at higher risk for the development of skin-related complications, especially skin erosions and less common skin infections, notably in cases when large (high profile) IPGs are utilized. The use of smaller IPGs could reduce the number of these biological as well as hardware-related complications and associated revision surgeries.

Indexed as

Complex regional pain syndromeFailed back surgery syndromeHardware-related complicationsSkin erosionSpinal cord stimulationVisual Analog Scale

Identifiers

PMID39640317
PMCPMC11618751

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-SA
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.