ArticleMedical physics2025
Dose distributions of proton therapy plans are robust against lowering the resolution of CTs combined with increasing noise.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Comparison of cone beam computed tomography post-processing methods for online adaptive proton therapy of prostate cancer.Physics and imaging in radiation oncology · 2025Article
- Dose distributions of proton therapy plans are robust against lowering the resolution of CTs combined with increasing noise.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
backgroundTreatment planning in radiation therapy (RT) is performed on image sets acquired with commercial x-ray computed tomography (CT) scanners. Considering an increased frequency of verification scans for adaptive RT and the advent of alternatives to x-ray CTs, there is a need to review the requirements for image sets used in RT planning. PURPOSE: This study aims to derive the required image quality (IQ) for the computation of the dose distribution in proton therapy (PT) regarding spatial resolution and the combination of spatial resolution and noise. The knowledge gained is used to explore the potential for dose reduction in tomography-guided PT.
methodsMathematical considerations indicate that the required spatial resolution for dose computation is on the scale of the set-up margins fed into the robust optimization. This hypothesis was tested by processing retrospectively 12 clinical PT cases, which reflect a variety of tumor localizations. Image sets were low-pass filtered and were made noisy in a generic manner. Dose distributions on the modified CT scans were computed with a Monte-Carlo dose engine. The similarity of these dose distributions with clinical ones was quantified with the gamma-index (1 mm/1%). The potential reduction of the x-ray exposure compared to the planning CT scan was estimated.
resultsDose distributions within the irradiated volume were robust against low-pass filtering of the CTs with kernels up to a full-width-at-half-maximum of 4 mm, that is, the gamma pass rate (1 mm/1%) was
conclusionsThe requirements on IQ in terms of spatial resolution in combination with noise for computing the dose in PT are clearly lower than the IQ of current clinical planning. The results apply, for example, to ultra-low dose x-ray CTs, proton CTs with coarse spatial detection, and attenuation images from the joint reconstruction of time-of-flight PET scans.