Evidence map›Paper›PMID 39604946›Full record

ArticleJournal of orthopaedic surgery and research2024

Biomechanical evaluation of a newly designed locking plate for opening wedge high tibial osteotomy: stress distribution and stability in the presence of lateral hinge fracture.

Hyung Jun Park, Dong Hun Suh, Hyongtaek Hong, Kyung-Wook Nha, Hyungsuh Kim, Kyoung-Tak Kang, Jae Gyoon Kim

Abstract read
In one paragraph

Article in Journal of orthopaedic surgery and research, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. [Application progress of customized steel plates in osteotomy and orthopedic treatment of knee osteoarthritis].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2025
    Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Hyung Jun Park *Department of Orthopedic Surgery, Korea University Ansan Hospital, Korea University College of Medicine, Ansan-Si, Gyeongki-Do, Republic of Korea.
Dong Hun Suh *Department of Orthopedic Surgery, Korea University Ansan Hospital, Korea University College of Medicine, Ansan-Si, Gyeongki-Do, Republic of Korea.
Hyongtaek HongSkyve R&D Lab, Seoul, Republic of Korea.
Kyung-Wook NhaDepartment of Orthopedic Surgery, Ilsanpaik Hospital, Inje University College of Medicine, Goyangsi, Republic of Korea.
Hyungsuh KimDepartment of Orthopedic Surgery, Ilsanpaik Hospital, Inje University College of Medicine, Goyangsi, Republic of Korea.
Kyoung-Tak Kang *Skyve R&D Lab, Seoul, Republic of Korea. tagi1024@gmail.com.
Jae Gyoon Kim *Department of Orthopedic Surgery, Korea University Ansan Hospital, Korea University College of Medicine, Ansan-Si, Gyeongki-Do, Republic of Korea. jgkim7437@gmail.com.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe study aimed to evaluate whether a new OhtoFix plate reduced stress around the D-hole compared with an old OhtoFix and TomoFix plate. The study also assessed whether the new OhtoFix plate had biomechanical stability in a lateral hinge fracture (LHF).

methodsA finite element model of the proximal tibia was developed using cross-sectional images from a 62-year-old Asian woman. The model was designed to simulate opening wedge high tibial osteotomy (OWHTO) using three types of locking plates: the newly designed locking plate (new OhtoFix plate), the prior version of the OhtoFix plate (old OhtoFix plate), and the TomoFix plate. The peak von Mises stress (PVMS) was analyzed around the D-hole and across the entire plate including the impact of different LHF types classified according to the Takeuchi classification.

resultsThe new OhtoFix plate significantly reduced stress around the D-hole compared with the old OhtoFix and TomoFix plates, with peak stresses of 189.5 MPa, 251.5 MPa, and 233.3 MPa, respectively. Despite this improvement, the new OhtoFix plate did not surpass the TomoFix in terms of overall stress distribution across the entire plate. Additionally, in cases of LHF, although peak stress remained at the D-hole in both the old OhtoFix and TomoFix plates, the peak stress shifted to the C-hole in the new plate.

conclusionsThe new OhtoFix plate improved stress distribution around the D-hole (even in the presence of a LHF) compared to the old OhtoFix plate and TomoFix plates. However, although the new OhtoFix plate reduced peak stress around the D-hole, it did not demonstrate superior overall stress distribution across the entire plate compared to the TomoFix plate.

Indexed as

Bone PlatesFinite Element AnalysisOsteotomyStress, MechanicalTibiaTibial FracturesBiomechanical PhenomenaFemaleHumansMiddle AgedLateral hinge fractureLocking plate biomechanicsOpening wedge high tibial osteotomyStress distribution

Identifiers

PMID39604946
PMCPMC11603658

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.