ArticleJournal of applied clinical medical physics2025
Development of dedicated phantom for end-to-end testing of multiple image-guidance configurations incorporating light-section-based surface guidance system.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Comparison of DVH-based machine learning and 3D convolutional neural network approaches for automated VMAT planning in head and neck cancer.Radiological physics and technology · 2026Article
- Impact of surface-guided prepositioning and respiratory coaching on the target localization accuracy in lung stereotactic body radiation therapy.Radiological physics and technology · 2026Article
- Impact of eFOV CT distortion on surface-guided radiotherapy registration: Phantom-based analysis.Journal of applied clinical medical physics · 2025Article
- Clinical technique: evaluation of detection accuracy in surface-guided radiation therapy using CT references from different treatment planning systems.Radiological physics and technology · 2025Article
- Development of dedicated phantom for end-to-end testing of multiple image-guidance configurations incorporating light-section-based surface guidance system.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeRadiotherapy devices with multiple image-guidance systems, such as surface-guided radiation therapy (SGRT), have been widely used in recent years. However, in the case of SGRT devices using the light-section method, coordinate coincidence evaluation using a phantom for SGRT devices with patterned light projection is not appropriate. Hence, this study aims to develop a dedicated phantom able to evaluate both the detection accuracy and coordinate coincidence of multiple IGRT configurations, including light-section-based SGRT devices. MATERIALS AND
methodFirst, we developed an end-to-end (E2E) phantom that can be scanned by CT and detected by a light-section-method-based SGRT device. Second, the detection accuracy of the phantom under three reference data acquisition conditions was evaluated using the E2E phantom. The three conditions were a body surface image detected by VOXELAN in the simulation CT room, a body surface image reconstructed from the volume data of the simulation CT room, and a body surface image acquired by VOXELAN in the radiotherapy room. Finally, the coordinate coincidence of the image-guidance system was evaluated using the E2E phantom.
resultUpon comparing detection accuracy among the three reference data acquisition methods, we found that the reference data generated in the CT room had the largest error (0.58 mm at maximum). The coordinate coincidences of the multiple image-guidance systems were within 1 mm for all components after the maintenance of VOXELAN using E2E phantom. Furthermore, the long-term direction stability was worse in the longitudinal direction.
conclusionThe new E2E phantom can be used to evaluate the detection accuracy of a light-section-based SGRT system and the coordinate coincidence using a Winston-Lutz-based method in multiple image-guided configurations. The detection accuracy in the three different reference images of VOXELAN using this phantom improved to within 1 mm in all directions.