ArticleMagnetic resonance in medicine2025
3D MR elastography at 0.55 T: Concomitant field effects and feasibility.
Article in Magnetic resonance in medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Single-Breathhold 3D MR Elastography in the Liver, With Simultaneous R2* and PDFF Mapping.Magnetic resonance in medicine · 2026Article
- Cartesian MaxGIRF: Model-based EPI reconstruction incorporating gradient nonlinearity and concomitant field effects.Magnetic resonance in medicine · 2026Article
- 3D MR elastography at 0.55 T: Concomitant field effects and feasibility.Magnetic resonance in medicine · 2025Article
- MR Elastography Using the Gravitational Transducer.Sensors (Basel, Switzerland) · 2024Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
purposeTo demonstrate the feasibility of hepatic 3D MR elastography (MRE) at 0.55 T in healthy volunteers using Hadamard encoding and to study the effects of concomitant fields in the domain of MRE in general.
methodsConcomitant field effects in MRE are assessed using a Taylor series expansion and an encoding scheme is proposed to study the corresponding effects on 3D MRE at 0.55 T in numerical simulations and in phantom experiments. In addition, five healthy volunteers were enrolled and scanned at 60 Hz mechanical excitation with a Hadamard-encoded 3D MRE sequence at 0.55 T and were also scanned with a reference 3D MRE sequence at 3 T for comparison. The retrieved biomechanical parameters were the magnitude of the complex shear modulus (|G*|), the shear wave speed (Cs), and the loss modulus (G″). Comparison of apparent SNR between 3 T and 0.55 T was performed.
resultsTheoretical analysis, numerical simulations and phantom experiments demonstrated that the effects of concomitant fields in 3D MRE at 0.55 T are negligible. In the healthy volunteer experiments, the mean values of |G*|, Cs, and G″ in the liver were 2.1 ± 0.3 kPa, 1.5 ± 0.1 m/s, and 0.8 ± 0.1 kPa at 0.55 T, respectively, and 2.0 ± 0.2 kPa, 1.5 ± 0.1 m/s, and 0.9 ± 0.1 kPa at 3 T, respectively. Bland-Altman analysis demonstrated good agreement between the biomechanical parameters retrieved at 0.55 T and 3 T. A 2.1-fold relative apparent SNR decrease was observed in 3D MRE at 0.55 T in comparison with 3 T.
conclusionHepatic 3D MRE is feasible at 0.55 T, showing promising initial results in healthy volunteers.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials