CFD investigations of a shape-memory polymer foam-based endovascular embolization device for the treatment of intracranial aneurysms.

Tanner L Cabaniss, Ryan Bodlak, Yingtao Liu, Geoffrey P Colby, Hyowon Lee, Bradley N Bohnstedt, Rinaldo Garziera, Gerhard A Holzapfel, Chung-Hao Lee

Abstract read
In one paragraph

Article in Biomechanics and modeling in mechanobiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

9 authors.

Tanner L Cabaniss *School of Aerospace and Mechanical Engineering, The University of Oklahoma, Norman, OK, 73019, USA.
Ryan Bodlak *School of Aerospace and Mechanical Engineering, The University of Oklahoma, Norman, OK, 73019, USA.
Yingtao LiuSchool of Aerospace and Mechanical Engineering, The University of Oklahoma, Norman, OK, 73019, USA.
Geoffrey P ColbyDepartment of Neurosurgery, UCLA Health, Los Angeles, CA, 90095, USA.
Hyowon LeeWeldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, 47907, USA.
Bradley N BohnstedtDepartment of Neurological Surgery, Indiana University School of Medicine, Indianapolis, IN, 46202, USA.
Rinaldo GarzieraDepartment of Engineering for Industrial Systems and Technologies, University of Parma, Parma, Italy.
Gerhard A HolzapfelInstitute of Biomechanics, Graz University of Technology, Graz, Austria.
Chung-Hao LeeBiomechanics and Biomaterials Design Laboratory (BBDL), Department of Bioengineering, University of California Riverside, Materials Science and Engineering (MS & E) Building, Room 207, 900 University Ave., Riverside, CA, 92521, USA. chunghao.lee@ucr.edu.

Funding

Tracking and Evaluation CoreU54GM104938 · NIGMS · UNIVERSITY OF OKLAHOMA HLTH SCIENCES CTR · PI TIMOTHY M VANWAGONER · 2013 to 2026
$68.2M
Improving outcomes in endovascular treatment of intracranial aneurysms: Combining additive manufacturing, in-silico modeling, and shape memory polymersR01HL159475 · NHLBI · UNIVERSITY OF OKLAHOMA · PI Chung-Hao Lee · 2022 to 2026
$3.3M
MUR-Italian Ministry of Universities and Research D.M. 737/2021-PNR-PNRR - NextGenerationEUNHLBI NIH HHS R01 HL159475NIGMS NIH HHS U54 GM104938NIH HHS R01HL159475Oklahoma Center for the Advancement of Science and Technology HR18-002Oklahoma Shared Translational and Clinical Resources NIGMS U54GM104938
6 · The paper itself

Abstract

The hemodynamic and convective heat transfer effects of a patient-specific endovascular therapeutic agent based on shape-memory polymer foam (SMPf) are evaluated using computational fluid dynamics studies for six patient-specific aneurysm geometries. The SMPf device is modeled as a continuous porous medium with full expansion for the flow studies and with various degrees of expansion for the heat transfer studies. The flow simulation parameters were qualitatively validated based on the existing literature. Further, a mesh independence study was conducted to verify an optimal cell size and reduce the computational costs. For convective heat transfer, a worst-case scenario is evaluated where the minimum volumetric flow rate is applied alongside the zero-flux boundary conditions. In the flow simulations, we found a reduction of the average intra-aneurysmal flow of > 85% and a reduction of the maximum intra-aneurysmal flow of > 45% for all presented geometries. These findings were compared with the literature on numerical simulations of hemodynamic and heat transfer of SMPf devices. The results obtained from this study provide a novel and practical framework for optimizing the design of patient-specific SMPf devices, integrating advanced computational models of hemodynamics and heat transfer. This framework could guide the future development of personalized endovascular embolization solutions for intracranial aneurysms with improved therapeutic outcome.

Indexed as

Embolization, TherapeuticEndovascular ProceduresHydrodynamicsIntracranial AneurysmPolymersSmart MaterialsComputer SimulationHemodynamicsHumansPorosityPolymersSmart MaterialsCerebral aneurysmsComputational fluid dynamicsIntracranial aneurysmsPatient-specific therapeuticsShape-memory polymers

Identifiers

PMID39585527
PMCPMC12090875

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.