CT radiation dose reduction with tin filter for localisation/characterisation level image quality in PET-CT: a phantom study.

Natalie Anne Bebbington, Lone Lange Østergård, Kenneth Boye Christensen, Paw Christian Holdgaard

Abstract read
In one paragraph

Article in EJNMMI physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. How can children benefit from total-body PET?The British journal of radiology · 2026
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Natalie Anne BebbingtonSiemens Healthcare A/S, Borupvang 9, Ballerup, 2750, Denmark. natalie.bebbington@siemens-healthineers.com.ORCID http://orcid.org/0000-0003-4833-9030
Lone Lange ØstergårdDepartment of Nuclear Medicine, Lillebaelt University Hospital, Beriderbakken 4, Vejle, 7100, Denmark.
Kenneth Boye ChristensenDepartment of Nuclear Medicine, Lillebaelt University Hospital, Beriderbakken 4, Vejle, 7100, Denmark.
Paw Christian HoldgaardDepartment of Nuclear Medicine, Lillebaelt University Hospital, Beriderbakken 4, Vejle, 7100, Denmark.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe tin filter has allowed radiation dose reduction in some standalone diagnostic computed tomography (CT) applications. Yet, 'low-dose' CT scans are commonly used in positron emission tomography (PET)-CT for lesion localisation/characterisation (L/C), with higher noise tolerated. Thus, dose reductions permissible with the tin filter at this image quality level may differ. The aim was to determine the level of CT dose reduction permitted with the tin filter in PET-CT, for comparable image quality to the clinical reference standard (CRS) L/C CT images acquired with standard filtration. MATERIALS AND

methodsA whole-body CT phantom was scanned with standard filtration in CRS protocols, using 120 kV with 20mAs-ref for bone L/C (used in

resultsQuantitative analysis demonstrated dose savings of 50-76% in bone, 27-51% in lung and 8-61% in soft tissue with use of the tin filter at Sn100kV. Qualitative analysis demonstrated dose reductions using Sn100kV in general agreement with the dose reductions indicated by quantitative analysis. Overall, CT dose reductions of around 85% were indicated for NaF bone PET-CT, allowing whole-body CT at just 0.2mSv ED, and a 30-40% CT dose reduction for FDG PET-CT using Sn100kV (1.7-2.0mSv), providing comparable image quality to current CRS images with standard filtration. Sn140kV demonstrated limited value in CT dose reduction.

conclusionsLarge CT dose reductions can be made using the tin filter at Sn100kV, when imaging bone, lung and soft tissue at L/C level CT image quality in PET-CT. As well as reducing the risk of inducing a cancer in later life, such dose reductions may also impact PET-CT practice, such as justifying cross-sectional over planar imaging or justifying PET-CT in younger patients.

Indexed as

CTDose reductionHybrid imagingPETSpectral shapingTin filter

Identifiers

PMID39585489
PMCPMC11589033