Accuracy of Axial Length Measurements by Two Swept-Source Optical Coherence Tomography Biometers in Macula-off Rhegmatogenous Retinal Detachment Eyes.
Article in Clinical ophthalmology (Auckland, N.Z.), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Optimizing axial length estimation for intraocular lens power calculation in phacovitrectomy for macula-off retinal detachment.International journal of retina and vitreous · 2025Article
- Studying the added effect of sum-of-segments biometry to modern intraocular lens power calculation formulas for short eyes.BMC ophthalmology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Purpose: To assess the accuracy of axial length (AXL) measurements using two swept-source optical coherence biometers, IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany) and ARGOS (Alcon, Inc. Fort Worth, TX), for macula-off rhegmatogenous retinal detachment (RRD). Methods: This retrospective study included 100 eyes with phakic primary macula-off RRD. Preoperative AXL measurements were performed using different methods: applanation A-scan ultrasound (U/S) biometry, combined applanation vector A/B-scan biometry, and optical biometry measurements were obtained using IOLMaster 700 (Carl Zeiss, Meditec, Jena, Germany) and ARGOS (Alcon, Inc. Fort Worth, TX). All patients underwent pars plana phacovitrectomy. At 8-10 weeks postoperatively, optical biometry was performed to record AXL. Results: Mean preoperative AXL measured using vector-A/B-scan ultrasonography was higher than that of postoperative AXL measured using IOLMaster (p < 0.05). Mean AXL measured by the standard mode of the ARGOS optical biometer was lower than mean AXL measured by both enhanced retina visualization (ERV) mode and user-adjusted method (p < 0.05). Mean same-eye AXL measured using IOLMaster was lower than that measured using ARGOS (p < 0.05). The least difference was observed with combined vector-A/B-scan ultrasound (on the positive side), followed by fellow eye AXL measured using IOLMaster optical biometry (on the negative side). Conclusion: Optical biometry of fellow eye in macula-off RRD was noted to be highly correlating with postoperative optical biometry of same eye using IOL Master 700 in eyes without anisometropia. IOLMaster 700 showed less accuracy in the AXL measurements for same eye. The ARGOS optical biometer may have a good potential for measuring same eye AXL. Using ERV mode or a user-adjusted method for the ARGOS optical biometer may improve accuracy of AXL measurements. Most accurate method for measuring AXL in same eye was vector-A/B-scan ultrasound.