ArticleAcute and critical care2024
Performance evaluation of non-invasive cardiac output monitoring device (HemoVista) based on multi-channel thoracic impedance plethysmography technology.
Article in Acute and critical care, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- New Insights into Cardiac Intensive Care.Reviews in cardiovascular medicine · 2026Review
- Wearable ECG-PPG Deep Learning Model for Cardiac Index-Based Noninvasive Cardiac Output Estimation in Cardiac Surgery Patients.Sensors (Basel, Switzerland) · 2026Article
- Noninvasive assessment of core metastatic genes in lung adenocarcinoma: development of a predictive model integrating single-cell transcriptomics and radiomics.Frontiers in oncology · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
Abstract
backgroundA non-invasive method of measuring cardiac output (CO) can be beneficial in the care of critically ill patients. HemoVista (BiLab Co., Ltd.) is a medical device that measures CO non-invasively using multi-channel impedance plethysmography technology. The purpose of this study was to exploratively evaluate the performance of HemoVista in critically ill patients undergoing CO monitoring with the FloTrac (Edwards Lifesciences).
methodsAfter non-invasively installing the HemoVista sensor in critically ill patients whose CO was monitored with the FloTrac, CO values measured by both devices were collected for 30 minutes. Cardiac output measured by both devices was selected every 10 seconds, creating approximately 360 data pairs per patient. Linear correlation analysis with Pearson correlation coefficients, Bland-Altman analysis, and four-quadrant plot analysis were performed to evaluate the performance of HemoVista.
resultsA total of 7,138 pairs of CO data from the 20 patients were included in the analysis. A significant correlation was observed between the two methods of measuring CO (Pearson's r=0.489, P<0.001). The mean bias was 1.03 L/min, the 95% CI for the limit of agreement was -1.83 L/min to 3.93 L/min and the percentage error was 55.8%. The concordance rate of time-dependent CO between the two devices was 14.6%.
conclusionsIt was observed that the current version of HemoVista has unsuitable performance for use in intensive care units. To be used for critically ill patients, the algorithm must be improved and reevaluated with an enhanced version.