ArticleEJNMMI physics2024
Partial volume correction for Lu-177-PSMA SPECT.
Article in EJNMMI physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 16 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
16 citing papers in PubMed.
- Influence of increased injected activities up 9.0 GBq [EJNMMI physics · 2026Article
- State-of-the-art and future developments in dosimetry for SSTR- and PSMA-targeting radioligand therapies.Zeitschrift fur medizinische Physik · 2026Review
- Development of energy deposition pixel kernel convolution for planar dosimetry inAnnals of nuclear medicine · 2026Article
- Investigation of iterative deconvolution inEJNMMI physics · 2026Article
- Direct estimation of activity concentration in regional voxels with application toMedical physics · 2026Article
- Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging.Medical physics · 2026Article
- Critical review of partial volume correction methods in PET and SPECT imaging: benefits, pitfalls, challenges, and future outlook.European journal of nuclear medicine and molecular imaging · 2026Review
- Revisiting the segmentation threshold for Lu-177 SPECT.Medical physics · 2026Article
- Towards practical radiopharmaceutical treatment planning: a review of dosimetry simplification techniques.Theranostics · 2026Review
- Quantitative SPECT/CT: Technique and Clinical Applications.Recent results in cancer research. Fortschritte der Krebsforschung. Progres dans les recherches sur le cancer · 2026Review
- MIRD Pamphlet No. 33: MIRDpvc-A Software Tool for Recovery Coefficient-Based Partial-Volume Correction.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2025Article
- Comparison of scatter and partial volume correction techniques for quantitative SPECT imaging ofEJNMMI physics · 2025Article
- Article
- Automatic bone marrow segmentation for precise [Medical physics · 2025Article
- 3D-Printed Organ-Realistic Phantoms to Verify Quantitative SPECT/CT Accuracy for 177Lu-PSMA-617 Treatment Planning.Pharmaceuticals (Basel, Switzerland) · 2025Article
- A review of state-of-the-art resolution improvement techniques in SPECT imaging.EJNMMI physics · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundThe limited spatial resolution in SPECT images leads to partial volume effect (PVE), degrading the subsequent dosimetric accuracy. We aim to quantitatively evaluate PVE and partial volume corrections (PVC), i.e., recovery coefficient (RC)-PVC (RC-PVC), reblurred Van-Cittert (RVC) and iterative Yang (IY), in
methodsWe employed a geometrical cylindrical phantom containing five spheres (diameters ranging from 20 to 40 mm) and 40 XCAT phantoms with various anatomical variations and activity distributions. SIMIND Monte Carlo code was used to generate realistic noisy projections. In the clinical study, sequential quantitative SPECT/CT imaging at 4 time-points post
resultsIn the simulation study, the spheres experienced significant PVE, i.e., 0.26 RC and 0.70 RC for the 20 mm and 40 mm spheres, respectively. RVC and IY improved the RC of the 20 mm sphere to 0.37 and 0.75 and RC of the 40 mm sphere to 0.96 and 1.04. Mismatch in mask increased the activity error for all spheres in IY. RVC increased noise and caused Gibbs ringing artifacts. For XCAT phantoms, both RVC and IY performed comparably and were superior to RC-PVC in reducing the MAE of the kidneys. However, IY and RC-PVC outperformed RVC for tumors. The XCAT phantom study and clinical study showed a similar trend in the kidney and tumor activity differences between non-PVC and PVC.
conclusionsPVE greatly impacts activity quantification, especially for small objects. All PVC methods improve the quantification accuracy in
Indexed as
Identifiers
What OpenQuestion holds
Registered trials