Evidence map›Paper›PMID 39516992›Full record

ArticleGeneral thoracic and cardiovascular surgery cases2024

Stent-induced new entry and device migration associated with hemodynamic stress after thoracic endovascular aortic repair for type B chronic aortic dissection using computational fluid dynamics analysis: a case report.

Itaru Hosaka, Takeshi Uzuka, Riko Umeta, Akihiko Sasaki

Abstract read
In one paragraph

Article in General thoracic and cardiovascular surgery cases, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Itaru HosakaDepartment of Cardiovascular surgery, Sunagawa City Medical Center, Nishi 4-jo Kita 3-chome 1-1, Sunagawa-shi, Hokkaido, Japan. itaruh@sapmed.ac.jp.ORCID http://orcid.org/0000-0002-4005-8229
Takeshi UzukaDepartment of Cardiovascular surgery, Sunagawa City Medical Center, Nishi 4-jo Kita 3-chome 1-1, Sunagawa-shi, Hokkaido, Japan.
Riko UmetaDepartment of Cardiovascular surgery, Sunagawa City Medical Center, Nishi 4-jo Kita 3-chome 1-1, Sunagawa-shi, Hokkaido, Japan.
Akihiko SasakiDepartment of Cardiovascular surgery, Sunagawa City Medical Center, Nishi 4-jo Kita 3-chome 1-1, Sunagawa-shi, Hokkaido, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundStent graft-induced new entry (SINE) after thoracic endovascular aortic repair (TEVAR) is a serious adverse event which leads to stent graft migration and rupture. SINE is known to be more frequent in cases of chronic dissection and oversizing. However, few studies have evaluated the influence of hemodynamic stress on SINE in patients with chronic dissection. Here, we report a rare case of TEVAR for chronic dissection with a dissection stent, inducing SINE 6 years after the first surgery. In addition, we analyze the hemodynamic stress for the aortic event using computational fluid dynamics (CFD) analysis. CASE PRESENTATION: A 69-year-old male underwent TEVAR with left subclavian artery debranching for chronic type B aortic dissection, using a stent graft and dissection stent. The postoperative course was uneventful, but follow-up computed tomography (CT) showed that the stent graft and bare-metal stent had dislodged 4 years after surgery. The gap between the bare-metal stent and the stent graft increased over time, and the proximal edge of the bare-metal stent led to SINE at the descending aorta 6 years after surgery. We performed reintervention to cover the SINE. The patient recovered well and was discharged at 6 days postoperatively. He is currently in good condition 6 months after reintervention. CFD analysis of the patient's CT image suggested that the local change in wall shear stress at the stent graft and dissection stent might be related to the aortic event.

conclusionHemodynamic stress is a factor affecting SINE and device migration. CFD may be useful for evaluating patient-specific risk of aortic events.

Indexed as

Computational fluid dynamicsDevice migrationStent-graft induced new entryThoracic endovascular aortic repairType B chronic aortic dissection

Identifiers

PMID39516992
PMCPMC11533579

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked