ArticleAnnals of biomedical engineering2025
The Effect of Low-Dose CT Protocols on Shoulder Model-Based Tracking accuracy Using Biplane Videoradiography.
Article in Annals of biomedical engineering, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Accuracy and Reliability of Synthetic Computed Tomography for Model-Based Tracking of Biplane Videoradiography Data.Annals of biomedical engineering · 2025Article
Corrections and comments
- Erratum issued
Authors and funding
5 authors.
Funding
Abstract
purposeModel-based tracking is being increasingly used to quantify shoulder kinematics and typically employs computed tomography (CT) to create the 3D bone volumes, which adds to the total radiation exposure. Lower-dose CT protocols may be possible given the contrast between bone and the surrounding soft tissues. The purpose of this study was to describe the dose-accuracy tradeoff between low-dose CT scans and the kinematic tracking accuracy of the humerus, scapula, and clavicle when tracked using an intensity-based registration algorithm.
methodsThree fresh-frozen cadavers consisting of the torso and bilateral shoulders were tested. The CT protocols investigated included a full-dose protocol and 4 experimental low-dose protocols that modulated x-ray tube current and peak voltage. Bead-based tracking (i.e., radiostereometric analysis) served as the reference standard to which model-based tracking results were compared. Accuracy was described in terms of both segmental (humerus, scapula, and clavicle) and joint (glenohumeral, acromioclavicular) kinematics using root-mean-square (RMSE), bias, precision, and worst-case errors.
resultsThe low-dose CT scans resulted in an average dose reduction of 70.6-92.8%. RMSEs tended to increase as CT dose decreased with average glenohumeral errors increasing from 0.5° and 0.6 mm to 0.6° and 0.6 mm between the highest and lowest-dose protocols, and average acromioclavicular errors increasing from 0.6° and 0.8 mm to 0.7° and 0.9 mm. However, the difference in joint kinematic errors between the highest and lowest-dose CT scanning protocols was generally small (≤0.3°, ≤ 0.1 mm).
conclusionIt is possible to substantially reduce the CT dose associated with shoulder motion analysis using biplane videoradiography without significantly impacting data fidelity.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.