ArticleArtificial organs2025
Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates.
Article in Artificial organs, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates.Artificial organs · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAn innovative solution has been recently proposed for the treatment of heart failure with preserved ejection fraction (HFpEF), using a centrifugal mechanical circulatory support (MCS) device. We sought firstly to assess the hemocompatibility of the proposed device. HFpEF treatment requires the blood pump to operate at low blood flow rate (0.05-0.5 L/min). Given high blood trauma expected in these conditions, we sought secondly to investigate design improvements in order to reduce this risk.
methodComputational fluid dynamics was used to analyze the blood fluid filled within the centrifugal pump and to estimate the blood trauma due to its operation. This assessment is based on the computation of integrated quantities such as the hemolysis index and the turbulent dissipation energy.
resultsThe hemolysis index associated with the present device is comparable to the figures from HVAD. With the proposed pump, for instance, an index of 0.0036 is obtained at 1 L/min and 3000 rpm. Using thinner blades for the impeller allows a 12% reduction of the hemolysis index in average, while reducing its diameter leads to an index 2.8 folds lower at 0.5 L/min.
conclusionThe present investigation shows the promising hemocompatibility of our centrifugal pump. Concerns about very high hemolysis generated at low flow rates could be overcome by reducing the impeller diameter. Experimental validations are planned to support our findings.