What CPAP to use in the delivery room? Bench comparison of two methods to provide continuous positive airways pressure in neonates.
Article in BMJ paediatrics open, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Leak during term neonatal CPAP stabilisation: does resuscitation device design affect delivered PEEP - a bench study.BMJ paediatrics open · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundContinuous positive airway pressure (CPAP) is a recommended first-line therapy for infants with respiratory distress at birth. Resuscitation devices incorporating CPAP delivery can have significantly different imposed resistances affecting airway pressure stability and work of breathing.
aimTo compare CPAP performance of two resuscitation devices (Neopuff T-piece resuscitator and rPAP) in a neonatal lung model simulating spontaneous breathing effort at birth.
methodsThe parameters assessed were variation in delivered pressures (∆P), tidal volume (VT), inspiratory effort (model pressure respiratory muscle (PRM)) and work of breathing (WOB). Two data sequences were required with Neopuff and one with rPAP: (1) set PRM with changes in VT and (2) constant VT (preterm 6 mL, term 22 mL) with increased effort. Data were collected at CPAP settings of 5, 7 and 9 cmH
results2298 breaths were analysed (760 rPAP, 795 Neopuff constant VT, 743 Neopuff constant PRM). With CPAP at 9 cmH
conclusionrPAP had smaller pressure swings than Neopuff at all CPAP levels and was thus more pressure stable. WOB was higher with Neopuff when VT was held constant. VT reduced with Neopuff when respiratory effort was constant.