Evidence map›Paper›PMID 39472953›Full record

ArticleTrials2024

Assessment of Cutting-Balloon Angioplasty with Novel Bioabsorbable Polymer-Coated Everolimus-Eluting Stent in Treating Calcified Coronary Lesions Guided by Intravascular Ultrasound (CUPID Trial): study design and protocol.

Jihun Ahn, HyeYon Yu, Sangho Park, Jon Suh

Registry-linked trialAbstract readClinical Trial Protocol
In one paragraph

Article in Trials, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT06177808 (Assessment of Cutting-Balloon Angioplasty With Novel Bioabsorbable Polymer-Coated, Everolimus-Eluting Stent in the Treatment of Calcified Coronary Lesions Guided by Intravascular Ultrasound), which is not on this map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT06177808 naunknown statusnot on this map

Assessment of Cutting-Balloon Angioplasty With Novel Bioabsorbable Polymer-Coated, Everolimus-Eluting Stent in the Treatment of Calcified Coronary Lesions Guided by Intravascular Ultrasound

TypeinterventionalSponsorSoonchunhyang University HospitalRan2024 to 2025Enrolled250ConditionsCoronary Artery DiseaseArmsWolverine cutting balloon, Non-compliant balloon
3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Review
  2. Enhanced Drug Effect With Cutting Balloon Followed by Drug Coated Balloon in a Rabbit Model.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2025
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Jihun AhnDepartment of Internal Medicine, Daejeon Eulji Medical Center, Eulji University School of Medicine, Daejeon, Korea.
HyeYon YuDepartment of Nursing, College of Medicine, Soonchunhyang University, Asan, Korea.
Sangho ParkDepartment of Internal Medicine, Soonchunhyang University Cheonan Hospital, Soonchunhyang University College of Medicine, Cheonan, Korea.
Jon SuhDepartment of Internal Medicine, Soonchunhyang University Bucheon Hospital, Soonchunhyang University College of Medicine, Bucheon, Korea. immanuel@schmc.ac.kr.ORCID http://orcid.org/0000-0001-9593-8453

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundVarious devices and techniques have been used for plaque modification in the treatment of severe coronary artery calcification. This prospective, multicenter, randomized, open-label study aims to evaluate the safety and efficacy of cutting balloon angioplasty using a novel bioabsorbable polymer-coated everolimus-eluting coronary stent for treating various degrees of calcified coronary lesions.

methodsWe outline the trial design aimed at assessing whether the cutting balloon (Wolverine™) is non-inferior to the non-compliant balloon in treating patients with calcified lesions, encompassing both de novo and in-stent restenosis (ISR) lesions. We aim to enroll 250 patients who have undergone bioabsorbable polymer-coated everolimus-eluting coronary stent (Synergy™) implantation. The primary endpoint is the minimal stent cross-sectional area at the calcium site as determined by intravascular ultrasonography. The secondary endpoints include major adverse cardiac events and target lesion revascularization at 12 months, alongside procedural convenience and operator-centric parameters, such as the number of balloons used, procedure time, and total contrast medium volume used. DISCUSSION: In this study, we will evaluate the efficacy and safety of Wolverine™ and non-compliant balloon in patients with calcified coronary lesions and provide a rationale for which type of balloons will optimally modify calcium lesions. In addition, we will attempt to expand the indications of the cutting balloon for treating mild-to-severe calcified coronary lesions. As the scope of insurance coverage for cutting balloons remains limited in some countries, this study may provide evidence for extending insurance coverage to the treatment of de novo calcified and ISR lesions.

trial registrationClinicalTrials.gov NCT06177808. Registered on January 1, 2024.

Indexed as

Absorbable ImplantsAngioplasty, Balloon, CoronaryCoronary Artery DiseaseDrug-Eluting StentsEverolimusMulticenter Studies as TopicPolymersUltrasonography, InterventionalVascular CalcificationAgedCardiovascular AgentsFemaleHumansMaleMiddle AgedProspective StudiesCardiovascular AgentsEverolimusPolymersCalcificationCutting balloonPercutaneous coronary intervention

Identifiers

PMID39472953
PMCPMC11520680

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.