Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya.
Article in Frontiers in global women's health, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026Article
- Incidence and risk factors for stillbirth in Kenya, Mozambique and the Gambia: the PRECISE cohort study.The Lancet regional health. Africa · 2026Article
- Frequency of suspected anemia in Japanese children aged 1-5 years using noninvasive hemoglobin measurement and influence of iron-fortified growing-up formula intake.BMC pediatrics · 2026Observational
- Diagnostic Performance and Trending Ability of Continuous Non-Invasive Hemoglobin Monitoring During Elective Intracranial Neurosurgery with Invasive Arterial Monitoring: Influence of Anesthetic Technique.Diagnostics (Basel, Switzerland) · 2026Article
- Feasibility of a Noncontact Photoplethysmography-Based Mobile App for Noninvasive Hemoglobin Monitoring: Exploratory Observational Study.JMIR formative research · 2026Observational
- Evaluation of non-invasive hemoglobin measurement in blood donors at a tertiary care hospital, Bangkok, Thailand.PloS one · 2026Article
- Comparison of Masimo Rad-67 SpHb non-invasive hemoglobin monitoring device with complete blood count measurement for use in pregnancy: An observational multi-site cohort study.medRxiv : the preprint server for health sciences · 2025Article
- Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
Background: Point of care hemoglobin meters play key roles in increasing access to anemia screening in antenatal care especially in settings with limited access to laboratories. We aimed to determine the diagnostic accuracy of a non-invasive spot-check hemoglobin (SpHb) meter, Masimo Rad-67® Pulse CO-Oximeter®, in the diagnosis of anemia in pregnant women attending antenatal care clinics in Kilifi, Kenya. Methods: This was a diagnostic accuracy study that retrospectively evaluated SpHb against a validated reference standard of laboratory assessed hemoglobin (Lab Hb) by a SYSMEX XN-330 automated hematology analyzer. The study was nested within a prospective pregnancy cohort study that recruited unselected pregnant women from antenatal care clinics in two public hospitals in Kilifi County, coastal Kenya. Records with both SpHb and Lab Hb were selected from pregnancy visits between May 2021 and December 2022. Linear regression and Bland-Altman analysis were performed to compare the two tests and diagnostic accuracy parameters obtained for the diagnosis of anemia. Results: A total of 2,975 records (from 2,203 unique participants), with paired SpHb and Lab Hb were analyzed. Linear regression showed a significant but weak positive correlation, a proportional bias of 0.44 (95% CI 0.41-0.47) and a constant of 7.59 (95% CI 7.30-7.87, Conclusion: Overall, SpHb by Masimo Rad-67® Pulse CO-Oximeter® did not accurately identify pregnant women with anemia and many cases would be missed. We would not recommend its use in antenatal care settings.