ArticleMagnetic resonance in medicine2025

3D distortion-free, reduced FOV diffusion-prepared gradient echo at 3 T.

Sarah McElroy, Raphael Tomi-Tricot, Jon Cleary, Hsern Ern Ivan Tan, Shawna Kinsella, Sami Jeljeli, Vicky Goh, Radhouene Neji

Abstract read
In one paragraph

Article in Magnetic resonance in medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Combined ADC and TMagnetic resonance in medicine · 2026
    Article
  2. Review
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Sarah McElroySchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.ORCID 0000-0002-4271-5931
Raphael Tomi-TricotSchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Jon ClearySchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Hsern Ern Ivan TanGuy's and St Thomas' NHS Foundation Trust, London, UK.ORCID 0000-0002-6945-3900
Shawna KinsellaSchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Sami JeljeliSchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Vicky GohSchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Radhouene NejiSchool of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo develop a 3D distortion-free reduced-FOV diffusion-prepared gradient-echo sequence and demonstrate its application in vivo for diffusion imaging of the spinal cord in healthy volunteers.

methodsA 3D multi-shot reduced-FOV diffusion-prepared gradient-echo acquisition is achieved using a slice-selective tip-down pulse in the phase-encoding direction in the diffusion preparation, combined with magnitude stabilizers, centric k-space encoding, and 2D phase navigators to correct for intershot phase errors. The accuracy of the ADC values obtained using the proposed approach was evaluated in a diffusion phantom and compared to the tabulated reference ADC values and to the ADC values obtained using a standard spin echo diffusion-weighted single-shot EPI sequence (DW-SS-EPI). Five healthy volunteers were scanned at 3 T using the proposed sequence, DW-SS-EPI, and a clinical diffusion-weighted multi-shot readout-segmented EPI sequence (RESOLVE) for cervical spinal cord imaging. Image quality, perceived SNR, and image distortion were assessed by two expert radiologists. ADC maps were calculated, and ADC values obtained with the proposed sequence were compared to those obtained using DW-SS-EPI and RESOLVE.

resultsConsistent ADC estimates were measured in the diffusion phantom with the proposed sequence and the conventional DW-SS-EPI sequence, and the ADC values were in close agreement with the reference values provided by the manufacturer of the phantom. In vivo, the proposed sequence demonstrated improved image quality, improved perceived SNR, and reduced perceived distortion compared to DW-SS-EPI, whereas all measures were comparable against RESOLVE. There were no significant differences in ADC values estimated in vivo for each of the sequences.

conclusion3D distortion-free diffusion-prepared imaging can be achieved using the proposed sequence.

Indexed as

Diffusion Magnetic Resonance ImagingEcho-Planar ImagingImage EnhancementImage Interpretation, Computer-AssistedImaging, Three-DimensionalSpinal CordAdultAlgorithmsFemaleHealthy VolunteersHumansMalePhantoms, ImagingReproducibility of Resultsdiffusion imagingdiffusion‐prepared imagingdistortion‐free diffusionreduced FOV diffusionspinal cord DWI