Evidence map›Paper›PMID 39400659›Full record

ArticleEuropean radiology experimental2024

Performance of dual-energy subtraction in contrast-enhanced mammography for three different manufacturers: a phantom study.

Gisella Gennaro, Giulia Vatteroni, Daniela Bernardi, Francesca Caumo

Abstract readComparative Study
In one paragraph

Article in European radiology experimental, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Gisella GennaroVeneto Institute of Oncology (IOV), IRCCS, Padua, Italy. gisella.gennaro@iov.veneto.it.ORCID http://orcid.org/0000-0003-2444-1778
Giulia VatteroniDepartment of Biomedical Sciences, Humanitas University, Milan, Italy.
Daniela BernardiDepartment of Biomedical Sciences, Humanitas University, Milan, Italy.
Francesca CaumoVeneto Institute of Oncology (IOV), IRCCS, Padua, Italy.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDual-energy subtraction (DES) imaging is critical in contrast-enhanced mammography (CEM), as the recombination of low-energy (LE) and high-energy (HE) images produces contrast enhancement while reducing anatomical noise. The study's purpose was to compare the performance of the DES algorithm among three different CEM systems using a commercial phantom.

methodsA CIRS Model 022 phantom, designed for CEM, was acquired using all available automatic exposure modes (AECs) with three CEM systems from three different manufacturers (CEM1, CEM2, and CEM3). Three studies were acquired for each system/AEC mode to measure both radiation dose and image quality metrics, including estimation of measurement error. The mean glandular dose (MGD) calculated over the three acquisitions was used as the dosimetry index, while contrast-to-noise ratio (CNR) was obtained from LE and HE images and DES images and used as an image quality metric.

resultsOn average, the CNR of LE images of CEM1 was 2.3 times higher than that of CEM2 and 2.7 times higher than that of CEM3. For HE images, the CNR of CEM1 was 2.7 and 3.5 times higher than that of CEM2 and CEM3, respectively. The CNR remained predominantly higher for CEM1 even when measured from DES images, followed by CEM2 and then CEM3. CEM1 delivered the lowest MGD (2.34 ± 0.03 mGy), followed by CEM3 (2.53 ± 0.02 mGy) in default AEC mode, and CEM2 (3.50 ± 0.05 mGy). The doses of CEM2 and CEM3 increased by 49.6% and 8.0% compared with CEM1, respectively.

conclusionOne system outperformed others in DES algorithms, providing higher CNR at lower doses. RELEVANCE STATEMENT: This phantom study highlighted the variability in performance among the DES algorithms used by different CEM systems, showing that these differences can be translated in terms of variations in contrast enhancement and radiation dose. KEY POINTS: DES images, obtained by recombining LE and HE images, have a major role in CEM. Differences in radiation dose among CEM systems were between 8.0% and 49.6%. One DES algorithm achieved superior technical performance, providing higher CNR values at a lower radiation dose.

Indexed as

AlgorithmsContrast MediaMammographyPhantoms, ImagingFemaleHumansRadiation DosageRadiographic Image EnhancementSubtraction TechniqueContrast MediaContrast mediaMammographyPhantoms (imaging)Radiation dosageRadiographic image enhancement

Identifiers

PMID39400659
PMCPMC11473475

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.