Influence of compliance and resistance of the test lung on the accuracy of the tidal volume delivered by the ventilator.

Zheng-Long Chen, Yu-Zhong Yan, Hong-Yi Yu, Qiu-Bo Wang, Wei Wang, Ming Zhong

Abstract read
In one paragraph

Article in BMC pulmonary medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Zheng-Long ChenDepartment of Respiratory and Critical Care Medicine, Zhoupu Hospital, Shanghai University of Medicine and Health Sciences, 1500 Zhouyuan Road, Pudong New Area, Shanghai, 201318, China.
Yu-Zhong YanLaboratory Department, Zhoupu Hospital, Shanghai University of Medicine and Health Sciences, 1500 Zhouyuan Road, Pudong New Area, Shanghai, 201318, China.
Hong-Yi YuShanghai Institute of Medical Device Testing, 1 Jinyinhua Road, Pudong New Area, Shanghai, 201318, China.
Qiu-Bo WangDepartment of Respiratory and Critical Care Medicine, Zhoupu Hospital, Shanghai University of Medicine and Health Sciences, 1500 Zhouyuan Road, Pudong New Area, Shanghai, 201318, China.
Wei WangShanghai Institute of Medical Device Testing, 1 Jinyinhua Road, Pudong New Area, Shanghai, 201318, China.
Ming ZhongDepartment of Critical Care Medicine, Zhongshan Hospital, Fudan University, 180 Fenglin Road, Xuhui District, Shanghai, 200032, China. zhong.ming@zs-hospital.sh.cn.

Funding

Natural Science Foundation of Shanghai Municipality 21ZR1428300Program of Science and Technology of Shanghai Pudong New Area PKJ2023-Y12Research Grant for Pudong Health Bureau of Shanghai YC-2023-0401
6 · The paper itself

Abstract

backgroundLarge variations in respiratory system compliance and resistance may cause the accuracy of tidal volume (VT) delivery beyond the declared range. This study aimed at evaluating the accuracy of VT delivery using a test lung model to simulate pulmonary mechanics under normal or disease conditions.

methodsIn vitro assessment of the VT delivery accuracy was carried out on two commercial ventilators. Measurements of the inspired and expired VT from the ventilator and FlowAnalyser were compared to evaluate the separated and combined influences of compliance and resistance on the delivered VT accuracy. To do this, the errors of five delivered volumes (30 ml, 50 ml, 100 ml, 300 ml, and 500 ml) were checked under 29 test conditions involving a total of 27 combinations of resistance and compliance.

resultsFor the tested ventilator S1 with a flow sensor near the expiratory valve, the average of expired VT errors (ΔVTexp) in three measurements (4 test conditions for each measurement) correlated to test lung compliance (r=-0.96, p = 0.044), and the average of inspired VT errors (ΔVTins) correlated to compliance (r = 0.89, p = 0.106); for the tested ventilator S2 with a flow sensor located at the Y piece, no clear relationship between compliance and ΔVTexp or ΔVTins was found. Furthermore, on two ventilators tested, the current measurements revealed a poor correlation between test lung resistance and ΔVTins or ΔVTexp, and the maximum values of ΔVTexp and ΔVTins correspond to the maximum resistance of 200 cmH

conclusionsDepending on the placement site of the flow sensor in the ventilator circuit, the compliance and resistance of the test lung have different influences on the accuracy of VT delivery, which is further attributed to different fluid dynamics effects of the compliance and resistance. The main influence of compliance is to raise the peak inspiratory pressure Ppeak, thereby increasing the compression volume within the ventilator circuit; whereas a high resistance not only contributes to elevating Ppeak, but more importantly, it governs the gas flow conditions. Ppeak is a critical predictive indicator for the accuracy of the VT delivered by a ventilator.

Indexed as

LungTidal VolumeVentilators, MechanicalAirway ResistanceEquipment DesignHumansLung ComplianceRespiration, ArtificialRespiratory MechanicsAirway resistanceInternational standardLung complianceQuality testingTidal volumeVentilator

Identifiers

PMID39385191
PMCPMC11465742

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.