Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum.
Article in BMC biotechnology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
backgroundHomocysteine (HCY) is a sulfur-containing amino acid that is an independent or important risk factor for the occurrence of many chronic diseases and is one of the most important indicators for determining health risks. However, existing HCY detection methods do not meet the requirements of clinical diagnosis. Therefore, there is an urgent need to establish new detection methods to meet the needs of clinical detection.
resultsIn this study, we used the principle of competitive method to establish a new method for the determination of HCY in human serum using a chemiluminescent enzyme immunoassay in conjunction with a chemiluminescent assay instrument that uses magnetic microparticles as the solid phase of the immunoreaction. The established method achieved satisfactory results in terms of minimum detection limit, specificity, accuracy, and clinical application. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 1.94-5.05%, the inter-assay CV was 2.29-6.88%, and the recovery rate was 88.60-93.27%. Cross-reactivity with L-cysteine ranged from 0.0100 to 0.0200 μmol/L, and cross-reactivity with glutathione ranged from 0.0100 to 0.200 μmol/L, all of which were less than the limit of detection (LoD) of this method. The linear factor R of this method was greater than 0.99.
conclusionsIn summary, the developed method showed a good correlation with the product from Abbott. A total of 996 clinical patients with cardiovascular diseases were evaluated using the method developed in this study.