ArticleJapanese journal of radiology2025
Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study.
Article in Japanese journal of radiology, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Comparison of iodine quantification accuracy on prototype deep silicon photon-counting and energy-integrating detector CT.Abdominal radiology (New York) · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
purposeTo investigate the feasibility and accuracy of iodine quantification using PCD-CT in standard-pitch and high-pitch scanning at different scan parameters in a phantom model. MATERIALS AND
methodsFour inserts with known iodine concentrations (2, 5, 10, and 15 mg/mL) were placed in the removable CT phantom and scanned using high-pitch (3.2) and standard-pitch (0.8) modes on PCD-CT. Two tube voltages (120 and 140 kVp) and four radiation doses (1, 3, 5, and 10 mGy) were alternated. Each scan setting was repeated three times. Mean iodine concentration for each insert across three consecutive slices was recorded. Percentage absolute bias (PAB) was assessed for iodine quantification.
resultsA total of 96 acquisitions were conducted. In small phantom, the average for PAB was 2.96% (range: 1.75% ~ 4.56%) and 1.67% (range: 1.00% ~ 3.42%) for high-pitch and standard-pitch acquisitions, respectively. In large phantom, it was 3.72% (range: 1.75% ~ 5.97%) and 2.94% (range: 1.75% ~ 4.70%). Linear regression analysis revealed that only phantom size significantly influenced (P < 0.001) the accuracy of iodine quantification.
conclusionThe high-pitch scan mode in PCD-CT can be used to quantify iodine density with similar accuracy compared with standard pitch.
Indexed as
Identifiers
39382795What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.