Evidence map›Paper›PMID 39380050›Full record

ArticleCritical care (London, England)2024

Blood trauma in veno-venous extracorporeal membrane oxygenation: low pump pressures and low circuit resistance matter.

Christopher Blum, Micha Landoll, Stephan E Strassmann, Ulrich Steinseifer, Michael Neidlin, Christian Karagiannidis

Abstract read
In one paragraph

Article in Critical care (London, England), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed.

  1. Extracorporeal COCritical care (London, England) · 2026
    Trial
  2. Article
  3. Article
  4. Haemolytic injury and cardiovascular disease.Nature reviews. Cardiology · 2026
    Review
  5. Protruding Objects in the Membrane Lung Outlet May Increase Thrombogenicity: Fluid Dynamical Insights.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
    Article
  6. Article
  7. [Design of a benchmark pump model and optimization of hemolysis testing protocol for evaluation of blood pump hemocompatibility].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2026
    Article
  8. Frontiers in pediatrics · 2025
    Article
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Christopher BlumDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany. blum@ame.rwth-aachen.de.
Micha LandollDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.
Stephan E StrassmannDepartment of Pneumology and Critical Care Medicine, ARDS and ECMO Centre, Cologne-Merheim Hospital, Cologne, Germany.
Ulrich SteinseiferDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.
Michael Neidlin *Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.
Christian Karagiannidis *Department of Pneumology and Critical Care Medicine, ARDS and ECMO Centre, Cologne-Merheim Hospital, Cologne, Germany.

Funding

Deutsche Forschungsgemeinschaft 467133626
6 · The paper itself

Abstract

backgroundVeno-venous extracorporeal membrane oxygenation (VV ECMO) has become standard of care in patients with the most severe forms of acute respiratory distress syndrome. However, hemolysis and bleeding are one of the most frequent side effects, affecting mortality. Despite the widespread use of VV ECMO, current protocols lack detailed, in-vivo data-based recommendations for safe ECMO pump operating conditions. This study aims to comprehensively analyze the impact of VV ECMO pump operating conditions on hemolysis by combining in-silico modeling and clinical data analysis.

methodsWe combined data from 580 patients treated with VV ECMO in conjunction with numerical predictions of hemolysis using computational fluid dynamics and reduced order modeling of the Rotaflow (Getinge) and DP3 (Xenios) pumps. Blood trauma parameters across 94,779 pump operating points were associated with numerical predictions of shear induced hemolysis.

resultsMinimal hemolysis was observed at low pump pressures and low circuit resistance across all flow rates, whereas high pump pressures and circuit resistance consistently precipitated substantial hemolysis, irrespective of flow rate. However, the lower the flow rate, the more pronounced the influence of circuit resistance on hemolysis became. Numerical models validated against clinical data demonstrated a strong association (Spearman's r = 0.8) between simulated and observed hemolysis, irrespective of the pump type.

conclusionsIntegrating in-silico predictions with clinical data provided a novel approach in understanding and potentially reducing blood trauma in VV ECMO. This study further demonstrated that a key factor in lowering side effects of ECMO support is the maintenance of low circuit resistance, including oxygenators with the lowest possible resistance, the shortest feasible circuit tubing, and cannulae with an optimal diameter.

Indexed as

Extracorporeal Membrane OxygenationHemolysisAdultComputer SimulationFemaleHumansMaleMiddle AgedPressureRespiratory Distress SyndromeBlood pumpHemolysisMortalityVV ECMO

Identifiers

PMID39380050
PMCPMC11462722

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.