Evidence map›Paper›PMID 39374243›Full record

ArticleJournal of applied clinical medical physics2025

Long-term performance monitoring of a-Si 1200 electronic portal imaging device for dosimetric applications.

Ivan Kutuzov, Ryan Rivest, Eric VanUytven, Boyd McCurdy

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Ivan KutuzovDepartment of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.ORCID https://orcid.org/0009-0008-2539-1251
Ryan RivestDepartment of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.
Eric VanUytvenDepartment of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.
Boyd McCurdyDepartment of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeRecently, dosimetri applications of the electronic portal imaging device (EPID) in radiotherapy have gained popularity. Confidence in the robust and reliable dosimetric performance of EPID detectors is essential for their clinical use. This study aimed to evaluate the dosimetric performance of the a-Si 1200 EPID and assess the long-term stability of its response.

methodsWeekly measurements were performed on two clinically used TrueBeam linear accelerators (linacs) equipped with a-Si 1200 EPID detectors over a 2-year period. They included dark and flood calibration fields, and EPID response to an open field corrected for the long-term machine output drift measured with the secondary absolute dosimeters: an ion chamber and an ion chamber array. All measurements were performed using five photon beam energies and two imaging modes: continuous and dosimetry. The measurements were analyzed for constancy and the presence of long-term trends. Comparisons were made between the two linacs for each beam energy. Pixel sensitivity matrices (PSM) were determined semi-annually and analyzed for long-term constancy for both treatment machines.

resultsThe long-term variation of the dark and flood field signals, integrated across the EPID plane, over the entire observation period did not exceed 0.17% and 0.79%, respectively. The output-corrected EPID response showed long-term variation from 0.28% to 0.36%, depending on beam energy, while the short-term variation was 0.04%-0.07% for EPID and 0.02%-0.06% for secondary dosimeters. The long-term variation of secondary dosimeters was 0.2%-0.3%. PSMs were found to be stable to within 1% for 97.8% of pixels and 2% for 100% of pixels.

conclusionTechniques to monitor and assess the long-term performance of the a-Si 1200 EPID as a dosimeter were developed and implemented using two TrueBeam linacs. The long-term variation of the EPID response was within clinical tolerance indicated in AAPM TG-142 report, and the detector was shown to be stable and reproducible for routine clinical dosimetry.

Indexed as

Particle AcceleratorsRadiometryRadiotherapy DosageRadiotherapy Planning, Computer-AssistedCalibrationElectrical Equipment and SuppliesHumansRadiotherapy, Intensity-ModulatedEPIDEPID dosimetryEPID response stabilitylinac outputlinac QAmachine QARadiotherapy

Identifiers

PMID39374243
PMCPMC11713653

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.