Clinical implementation and patient-specific quality assurance solutions for real-time target tracking and dynamic delivery in Radixact synchrony.
Article in Journal of applied clinical medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Retrospective log-based quality assurance of real-time MLC adaptation on a helical tomotherapy platform.Journal of applied clinical medical physics · 2026Article
- Markerless tumor tracking for lung SBRT with the Radixact Synchrony system: Initial experience and short-term clinical outcomes.Journal of applied clinical medical physics · 2025Article
- Evaluation of dose delivery accuracy of the lung with the respiratory technique of the Radixact Synchrony real-time tumor tracking system.Technical innovations & patient support in radiation oncology · 2025Article
- Clinical implementation and patient-specific quality assurance solutions for real-time target tracking and dynamic delivery in Radixact synchrony.Journal of applied clinical medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe installation and testing of the first Radixact with Synchrony system in Colombia marked a significant milestone in Latin America's medical landscape. There was a need to devise a robust quality assurance protocol to comprehensively evaluate both dose delivery and motion tracking accuracy. However, testing experiences under clinical conditions have not been extensively reported. Additionally, there are limited recommended measuring devices for Synchrony evaluation. PURPOSE: To validate and implement an alternative setup for dynamic-PSQA while testing Synchrony's functionality under clinical scenarios, including real-patient motion traces, and to provide guidance to new centers undergoing clinical implementation of Helical Synchrony.
methodsThis approach involves using the Iba miniPhantomR with strategically placed fiducial markers for configuring Gafchromic-films and array-based setups. When paired with the CIRS Dynamic Platform, this enables an innovative dynamic setup with trackable features for Synchrony delivery testing. Assessment scenarios, including compensation (M1S1) and no-motion compensation (M1S0), were evaluated using 2D-gamma pass rate analysis with multiple clinical gamma criteria. The Synchrony-Simulation feature was used to assess pre-treatment performance and capture the patient's target motion pattern. Synchrony for common clinical cases with patient's motion-traces was validated.
resultsThe results for M1S0 and M1S1 demonstrated consistency with previous studies evaluating Synchrony functionality. Analysis using different gamma criteria unveiled dosimetric differences and impacts across various motion ranges. The application of effective kV-dose subtraction for array-based methods is of upmost importance when evaluating dynamic-PSQA with stringent gamma-criteria. However, no significant kV-dose impact on EBT3-Film was detectable.
conclusionTwo implemented configurations for dynamic-PSQA setups were validated and successfully integrated into our clinic. We addressed both the benefits and limitations of array-based and film-based methods. The functionality and limitations of Synchrony were evaluated using the proposed setups. The potential utility of Synchrony-Simulation, along with the proposed patient-case classification table, can offer valuable support for new users during the clinical implementation of Synchrony treatments.