Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study.
Article in PLOS global public health, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05257564 (Performance and Usability Evaluation of Cardiometabolic Point-of-care Devices in a Target Use Setting), which is not on this map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Performance and Usability Evaluation of Cardiometabolic Point-of-care Devices in a Target Use Setting
Who cites it
4 citing papers in PubMed.
- Analytical performance of four HbA1c point-of-care devices: a cross-sectional evaluation in low- and middle-income countries.BMJ open · 2026Trial
- Bridging the diagnostic gap in diabetes.Bulletin of the World Health Organization · 2026Article
- Implementation of multiparameter point-of-care testing devices for non-communicable diseases at primary healthcare level in Peru.BMC health services research · 2025Article
- Advancing clinical biochemistry: addressing gaps and driving future innovations.Frontiers in medicine · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Non-communicable diseases (NCDs), such as cardiovascular disease and diabetes, represent a serious global health concern. There is an urgent need for prompt diagnosis and effective monitoring at point of care, especially in low- and middle-income countries. Here we present the results of a study assessing the quantitative accuracy of two devices that may fit the target product profile for a cardiometabolic point-of-care device. This prospective, quantitative, accuracy study (NCT05257564) was conducted between March to May 2022, investigating the performance of the JanaCare Aina Blood Monitoring System (JCAina) and the Tascom SimplexTAS 101 device (TAS101) compared with local standard laboratory methods in rural Nepal. Using fingerstick capillary blood, cardiometabolic parameters were analysed using both devices. The quantitative accuracy was compared against a local laboratory reference assay. System usability was also assessed. For JCAina, the mean absolute biases (Bland-Altman analysis) for glucose, HbA1c and total cholesterol tests were -3.87 mg/dL (95% CI: -7.52--0.22), 1.34% (95% CI: 1.21-1.47), and -9.52 mg/dL (95% CI: -11.9--7.2), respectively, corresponding to mean percentage biases of 2.0%, 18.5%, and -6.4%. These indicate clinically small (<10% biases) differences from laboratory results for glucose and cholesterol, and a moderate (10-20%) positive bias for HbA1c. For TAS101, the mean absolute biases for glucose, HbA1c, total cholesterol and creatinine tests were 18.7 mg/dL (95% CI: 15.8-21.5), -0.2% (95% CI: -0.26--0.14), 29.8 mg/dL (95% CI: 27.0-32.6), and -0.02 mg/dL (95% CI: -0.05-0.01), respectively, corresponding to mean percentage biases of 12.1%, -2.6%, 15.8%, and -4.5%. These indicate clinically small differences for HbA1c and creatinine, and moderate positive biases for glucose and cholesterol. Both systems exhibited usability challenges. The JCAina and TAS101 point-of-care cardiometabolic devices were shown to have promising accuracy in environmental conditions such as in Nepal, though improvements are still needed for some parameters and for ease of use. Trial registration: NCT05257564 (ClinicalTrials.gov).