ArticleArchives of orthopaedic and trauma surgery2024
Adverse events related to robotic-assisted knee arthroplasty: a cross-sectional study from the MAUDE database.
Article in Archives of orthopaedic and trauma surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Robot-assisted vs. navigation-assisted total knee arthroplasty: A systematic review and meta-analysis of the latest controlled studies based on perioperative, functional, and radiographic outcomes.Journal of robotic surgery · 2026Pooled it
- Failure modes of robotic-assisted arthroplasty platforms in Australia: a descriptive analysis of TGA adverse event reports.Journal of robotic surgery · 2026Article
- Medicolegal aspects of robotic-assisted surgery: litigation trends, complication risks, and human factors interventions.Journal of robotic surgery · 2026Review
- Robotic-assisted and AI-augmented arthroplasty: converging technologies in joint replacement.Journal of robotic surgery · 2026Review
- Use of the VELYS Robotic-Assisted Solution in knee arthroplasty: A scoping review.Journal of experimental orthopaedics · 2026Review
- Global trends and hotspots in robot-assisted arthroplasty: a CiteSpace-based bibliometric and visualized analysis.Journal of robotic surgery · 2025Review
- Merits and Limitations of Robotic-assisted Surgery in Improving Precision, Accuracy, and Patient Outcomes in Orthopedic Procedures.Journal of orthopaedics and sports medicine · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
Abstract
backgroundRobotic-assisted surgical technique has been clinically available for decades, yet real-world adverse events (AEs) and complications associated with primary knee arthroplasty remain unclear.
methodsIn March 2023, we searched the FDA website and extracted AEs related to robotic assisted knee arthroplasty (RAKA) from the MAUDE database over the past 10 years. The "Brand Name" function queried major robotic platforms, including active and semi-active systems. The overall incidence of AEs was estimated based on annual surgical volume from the current American Joint Replacement Registry (AJRR). Two authors independently collected data on event date, event type, device problem, and patient problem.
resultsOf 839 eligible reports, device malfunction comprised mechanical failure (343/839, 40.88%) and software failure (261/839, 31.11%). For surgical complications, inappropriate bone resection (115/839, 13.71%) was most frequent, followed by bone/soft tissue damage (83/839, 9.89%). Notably, over-resection exceeding 2 mm (88/839, 10.49%), joint infection (25/839, 2.98%), and aseptic loosening (1/839, 0.12%) were major complications. Only two track pins related AEs were found. Moreover, the distribution of these AEs differed substantially between robot manufacturers. According to the AEs volume and AJRR data, the overall incidences of AEs related to RAKAs were calculated with 0.83% (839/100,892) between November 2010 and March 2023.
conclusionOur analysis shows that while reported AEs might be increasing for RAKAs, the overall rate remains relatively low. Reassuringly, device malfunction was the most commonly AEs observed, with a minor impact on postoperative outcomes. Furthermore, our data provide a benchmark for patients, surgeons, and manufacturers to evaluate RAKA performance, though continued improvement in reducing serious AEs incidence is warranted.
Indexed as
Identifiers
39311943What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.