Development of novel ionization chambers for reference dosimetry in electron flash radiotherapy.

Kevin Liu, Shannon Holmes, Ahtesham Ullah Khan, Brian Hooten, Larry DeWerd, Emil Schüler, Sam Beddar

Abstract read
In one paragraph

Article in Medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.

0numbers the graph read from it
0cells of the map it votes in
13citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

13 citing papers in PubMed.

  1. Article
  2. Article
  3. Volumetric FLASH dosimetryPhysics in medicine and biology · 2026
    Article
  4. Article
  5. Article
  6. Investigating the oxygen dependence of FLASH-RT using electron paramagnetic resonance imaging.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
    Article
  7. Article
  8. Review
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Kevin LiuDepartment of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Shannon HolmesStandard Imaging Inc., Middleton, Wisconsin, USA.
Ahtesham Ullah KhanDepartment of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Brian HootenStandard Imaging Inc., Middleton, Wisconsin, USA.
Larry DeWerdDepartment of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Emil SchülerDepartment of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Sam BeddarDepartment of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

Funding

Tumor Evolution and Metastasis ProgramP30CA016672 · NCI · UNIVERSITY OF TX MD ANDERSON CAN CTR · PI Fabien Delerue · 1985 to 2026
$290.8M
Research Supplements to Promote Diversity in Health-Related Research: Preclinical optimization of ultra-high dose rate (FLASH) radiotherapy parameters for translational relevanceR01CA266673 · NCI · UNIVERSITY OF TX MD ANDERSON CAN CTR · PI SCHUELER, EMIL · 2022 to 2025
$3.5M
Water-Equivalent Plastic Scintillation Detectors for Small-Field RadiotherapyR44CA153824 · NCI · STANDARD IMAGING, INC. · PI BEDDAR, SAM, SCHMIDT, DANIEL G. · 2013 to 2014
$961k
FY22 SBIR TOPIC 434 PHASE I - NOVEL NIST TRACEABLE IONIZATION CHAMBER FOR FLASH DOSIMETRY.75N91022C00039 · NCI · STANDARD IMAGING, INC. · PI HOLMES, SHANNON · 2022 to 2022
$400k
Cancer Prevention and Research Institute of Texas RP210042National Institutes of Health: National Cancer Institute P30CA016672National Institutes of Health: National Cancer Institute R01CA266673National Institutes of Health - SBIR 434-75N91022C00039NCI NIH HHS 75N91022C00039NCI NIH HHS P30 CA016672NCI NIH HHS R01 CA266673NCI NIH HHS R44 CA153824
6 · The paper itself

Abstract

backgroundReference dosimetry in ultra-high dose rate (UHDR) beamlines is significantly hindered by limitations in conventional ionization chamber design. In particular, conventional chambers suffer from severe charge collection efficiency (CCE) degradation in high dose per pulse (DPP) beams. PURPOSE: The aim of this study was to optimize the design and performance of parallel plate ion chambers for use in UHDR dosimetry applications, and evaluate their potential as reference class chambers for calibration purposes. Three chamber designs were produced to determine the influence of the ion chamber response on electrode separation, field strength, and collection volume on the ion chamber response under UHDR and ultra-high dose per pulse (UHDPP) conditions.

methodsThree chambers were designed and produced: the A11-VAR (0.2-1.0 mm electrode gap, 20 mm diameter collector), the A11-TPP (0.3 mm electrode gap, 20 mm diameter collector), and the A30 (0.3 mm electrode gap, 5.4 mm diameter collector). The chambers underwent full characterization using an UHDR 9 MeV electron beam with individually varied beam parameters of pulse repetition frequency (PRF, 10-120 Hz), pulse width (PW, 0.5-4 µs), and pulse amplitude (0.01-9 Gy/pulse). The response of the ion chambers was evaluated as a function of the DPP, PRF, PW, dose rate, electric field strength, and electrode gap.

resultsThe chamber response was found to be dependent on DPP and PW, and these dependencies were mitigated with larger electric field strengths and smaller electrode spacing. At a constant electric field strength, we measured a larger CCE as a function of DPP for ion chambers with a smaller electrode gap in the A11-VAR. For ion chambers with identical electrode gap (A11-TPP and A30), higher electric field strengths were found to yield better CCE at higher DPP. A PW dependence was observed at low electric field strengths (500 V/mm) for DPP values ranging from 1 to 5 Gy at PWs ranging from 0.5 to 4 µs, but at electric field strengths of 1000 V/mm and higher, these effects become negligible.

conclusionThis study confirmed that the CCE of ion chambers depends strongly on the electrode spacing and the electric field strength, and also on the DPP and the PW of the UHDR beam. A significant finding of this study is that although chamber performance does depend on PW, the effect on the CCE becomes negligible with reduced electrode spacing and increased electric field. A CCE of ≥95% was achieved for DPPs of up to 5 Gy with no observable dependence on PW using the A30 chamber, while still achieving an acceptable performance in conventional dose rate beams, opening up the possibility for this type of chamber to be used as a reference class chamber for calibration purposes of electron FLASH beamlines.

Indexed as

ElectronsRadiometryCalibrationEquipment DesignRadiotherapy DosageReference StandardsdosimetryFLASHion chamber