ArticleBMC medical informatics and decision making2024
Real-world data to support post-market safety and performance of embolization coils: evidence generation from a medical device manufacturer and data institute partnership.
Article in BMC medical informatics and decision making, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- National adverse event reporting for medical devices in Israel, 2013-2024.Israel journal of health policy research · 2026Article
- Leveraging artificial intelligence in antibody-drug conjugate development: from target identification to clinical translation in oncology.NPJ precision oncology · 2025Review
- Biomechanical investigations on compression, expansion, and flexion of tubal occluders: a finite element analysis.Frontiers in bioengineering and biotechnology · 2025Article
- Challenges and dynamics in reporting medical device incidents: a qualitative study.Frontiers in health services · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundRecognizing the limitations of pre-market clinical data, regulatory authorities have embraced total product lifecycle management with post-market surveillance (PMS) data to assess medical device safety and performance. One method of proactive PMS involves the analysis of real-world data (RWD) through retrospective review of electronic health records (EHR). Because EHRs are patient-centered and focused on providing tools that clinicians use to determine care rather than collecting information on individual medical products, the process of transforming RWD into real-world evidence (RWE) can be laborious, particularly for medical devices with broad clinical use and extended clinical follow-up. This study describes a method to extract RWD from EHR to generate RWE on the safety and performance of embolization coils.
methodsThrough a partnership between a non-profit data institute and a medical device manufacturer, information on implantable embolization coils' use was extracted, linked, and analyzed from clinical data housed in an electronic data warehouse from the state of Indiana's largest health system. To evaluate the performance and safety of the embolization coils, technical success and safety were defined as per the Society of Interventional Radiology guidelines. A multi-prong strategy including electronic and manual review of unstructured (clinical chart notes) and structured data (International Classification of Disease codes), was developed to identify patients with relevant devices and extract data related to the endpoints.
resultsA total of 323 patients were identified as treated using Cook Medical Tornado, Nester, or MReye embolization coils between 1 January 2014 and 31 December 2018. Available clinical follow-up for these patients was 1127 ± 719 days. Indications for use, adverse events, and procedural success rates were identified via automated extraction of structured data along with review of available unstructured data. The overall technical success rate was 96.7%, and the safety events rate was 5.3% with 18 major adverse events in 17 patients. The calculated technical success and safety rates met pre-established performance goals (≥ 85% for technical success and ≤ 12% for safety), highlighting the relevance of this surveillance method.
conclusionsGenerating RWE from RWD requires careful planning and execution. The process described herein provided valuable longitudinal data for PMS of real-world device safety and performance. This cost-effective approach can be translated to other medical devices and similar RWD database systems.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.