Evidence map›Paper›PMID 39298303›Full record

ArticleIEEE transactions on ultrasonics, ferroelectrics, and frequency control2024

A 9-Fr Endovascular Therapy Transducer With an Acoustic Metamaterial Lens for Rapid Stroke Thrombectomy.

Phuong T Vu, Stephan Strassle Rojas, Caroline C Ott, Brooks D Lindsey

Abstract read
In one paragraph

Article in IEEE transactions on ultrasonics, ferroelectrics, and frequency control, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Review
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Phuong T Vu
Stephan Strassle Rojas
Caroline C Ott
Brooks D Lindsey

Funding

3D Multi-Functional Catheter-Based Imaging of Coronary Lesion Composition, Structure, and Hemodynamics in Intermediate StenosesR01EB031101 · NIBIB · GEORGIA INSTITUTE OF TECHNOLOGY · PI LINDSEY, BROOKS D · 2021 to 2024
$2.5M
NIBIB NIH HHS R01 EB031101
6 · The paper itself

Abstract

Large vessel occlusion (LVO) stroke, in which major cerebral arteries such as the internal carotid and middle cerebral arteries supplying the brain are occluded, is the most debilitating form of acute ischemic stroke (AIS). The current gold standard treatment for LVO stroke is mechanical thrombectomy; however, initial attempts to recanalize these large, proximal arteries supplying the brain fail in up to 75% of cases, leading to repeated passes that decrease the likelihood of success and affect patient outcomes. We report the design, fabrication, and testing of a mm forward-treating ultrasound (US) transducer with an acoustic metamaterial lens to dissolve blood clots recalcitrant to first-pass mechanical thrombectomy in LVO stroke. Due to the lens with microscale features, the device was able to produce a increase in peak negative pressure (PNP) (4.3 versus 1.8 MPa) and increase in blood clot dissolution rate ( versus mg/min) with 90% mass reduction after 30 min of treatment. In this small endovascular form factor, the acoustic metamaterial lens increased the acoustic output from the transducer while minimizing the US energy delivered to the surrounding areas outside of the treatment volume.

Indexed as

Endovascular ProceduresStrokeThrombectomyTransducersUltrasonic TherapyAnimalsEquipment DesignHumans

Identifiers

PMID39298303
PMCPMC11875980

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.