Assessment of thermochromic phantoms for characterizing microwave ablation devices.
Article in Medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
- Update of
Authors and funding
6 authors.
Funding
Abstract
background and purposeThermochromic gel phantoms provide a controlled medium for visual assessment of thermal ablation device performance. However, there are limited studies reporting on the comparative assessment of ablation profiles assessed in thermochromic gel phantoms against those in ex vivo tissue. The objective of this study was to compare microwave ablation zones in a thermochromic tissue-mimicking gel phantom and ex vivo bovine liver and to report on measurements of the temperature-dependent dielectric and thermal properties of the phantom.
methodsThermochromic polyacrylamide phantoms were fabricated following a previously reported protocol. Phantom samples were heated to temperatures in the range of 20°C-90°C in a temperature-controlled water bath, and colorimetric analysis of images of the phantom taken after heating was used to develop a calibration between color changes and the temperature to which the phantom was heated. Using a custom, 2.45 GHz water-cooled microwave ablation antenna, ablations were performed in fresh ex vivo liver and phantoms using 65 W applied for 5 min or 10 min (n = 3 samples in each medium for each power/time combination). Broadband (500 MHz-6 GHz) temperature-dependent dielectric and thermal properties of the phantom were measured over the temperature range of 22°C-100°C.
resultsColorimetric analysis showed that the sharp change in gel phantom color commences at a temperature of 57°C. Short and long axes of the ablation zone in the phantom (as assessed by the 57°C isotherm) for 65 W, 5 min ablations were aligned with the extents of the ablation zone observed in ex vivo bovine liver. However, for the 65 W, 10 min setting, ablations in the phantom were on average 23.7% smaller in the short axis and 7.4 % smaller in the long axis than those observed in ex vivo liver. Measurements of the temperature-dependent relative permittivity, thermal conductivity, and volumetric heat capacity of the phantom largely followed similar trends to published values for ex vivo liver tissue.
conclusionThermochromic tissue-mimicking phantoms provides a controlled, and reproducible medium for comparative assessment of microwave ablation devices and energy delivery settings. However, ablation zone size and shapes in the thermochromic phantom do not accurately represent ablation sizes and shapes observed in ex vivo liver tissue for high energy delivery treatments (65 W, 10 min). One cause for this limitation is the difference in temperature-dependent thermal and dielectric properties of the thermochromic phantom compared to ex vivo bovine liver tissue, as reported in the present study.