Evidence map›Paper›PMID 39286713›Full record

ArticleCureus2024

Utilizing Leadless Pacemakers in Extremely Elderly Patients With a Conventional Pacemaker System: A Two-Year Follow-Up Case Series Without Generator Extraction in High-Risk Scenarios.

Man Fong Chu, Weng Chio Tam, Kuok Wun Lam, Chon Hou Chan, Màrio Évora, U Po Lam

Abstract read
In one paragraph

Article in Cureus, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Man Fong ChuCardiology, Centro Hospitalar Conde São Januário, Macau, MAC.
Weng Chio TamCardiology, Centro Hospitalar Conde São Januário, Macau, MAC.
Kuok Wun LamCardiology, Centro Hospitalar Conde São Januário, Macau, MAC.
Chon Hou ChanDermatology, Centro Hospitalar Conde São Januário, Macau, MAC.
Màrio ÉvoraCardiology, Centro Hospitalar Conde São Januário, Macau, MAC.
U Po LamCardiology, Centro Hospitalar Conde São Januário, Macau, MAC.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background and objectives Leadless pacemakers, known for their safer clinical profile, offer significant advantages for elderly patients at a higher risk of complications associated with transvenous pacemaker procedures, particularly those susceptible to high-risk bleeding and infections related to cardiac implantable electronic device interventions. This study explores an alternative use of leadless pacemakers without removing existing transvenous systems, deviating from conventional generator replacement and lead re-interventions. Methods This study was conducted with full approval from the Institutional Review Board, Medical Ethical Committee, Centro Hospitalar Conde São Januário, Macau. Between January 2018 and December 2021, we conducted a retrospective case series involving extremely elderly individuals (aged 85 years or older) at a high risk of complications, necessitating either generator replacement or lead re-implantation. The study considered implanting a leadless pacemaker (Micra; Medtronic, Minneapolis, MN, USA) without removing the transvenous generator. For the primary endpoints, we evaluated procedure-related complications and clinical outcomes during hospitalization. Secondary endpoints included the stability of parameters and any unexpected interference or interactions between the two systems during the two-year follow-up. Results Eleven patients (aged 86-101) were enrolled, most receiving antiplatelet or anticoagulation therapy. Leadless pacemaker implantation proceeded without major complications or adverse clinical outcomes during hospitalization. Regular follow-up was conducted every three to six months for adjusting pacemaker parameters and interrogating each patient. Over two years, three patients died from non-cardiac causes: two from infection and one from spontaneous intracranial hemorrhage, while eight completed regular follow-ups. We didn't detect any episodes of ventricular arrhythmias or intracardiac capture from the transvenous pacemaker system. We observed the stability in both the longevity and the voltage of the conventional generator battery, maintaining similar parameters without significant depletion (mean voltage decline: -0.07V/year). Parameters of the leadless pacemaker remained consistently normal without interference with existing pacing systems. Conclusion Implanting leadless pacemakers without removing transvenous pacemaker generators appears safe and effective for extremely elderly patients who are at high risk of complications. Comprehensive two-year follow-up supports the safety and viability of this approach. Opting for this approach instead of conventional generator replacement, with or without additional lead implantation, may be reasonable in this population. However, further research within this patient cohort, such as exploring long-term outcomes beyond two years or comparing clinical outcomes with conventional strategies, may be necessary.

Indexed as

atrial fibrillationcardiac implantable electronic deviceend of serviceextreme elderlygenerator replacementleadless pacemaker

Identifiers

PMID39286713
PMCPMC11403653

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.