A full-custom fully implantable cochlear implant system validated in vivo with an animal model.

Hasan Uluşan, M Berat Yüksel, Özlem Topçu, H Andaç Yiğit, Akın M Yılmaz, Mert Doğan, Nagihan Gülhan Yasar, İbrahim Kuyumcu, Aykan Batu, Nebil Göksu and 2 more

Abstract read
In one paragraph

Article in Communications engineering, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Review
  2. Article
  3. Review
  4. Article
  5. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors.

Hasan Uluşan *Department of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.ORCID http://orcid.org/0000-0001-8788-8863
M Berat Yüksel *Department of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.
Özlem TopçuDepartment of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.
H Andaç YiğitDepartment of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.
Akın M YılmazDepartment of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.
Mert DoğanDepartment of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.ORCID http://orcid.org/0009-0001-5617-8372
Nagihan Gülhan YasarDepartment of Otorhinolaryngology, Faculty of Medicine, Gazi University, Ankara, Turkey.ORCID http://orcid.org/0000-0003-1207-3584
İbrahim KuyumcuDepartment of Otorhinolaryngology, Faculty of Medicine, Gazi University, Ankara, Turkey.
Aykan BatuMETU-MEMS Research and Applications Center, Ankara, Turkey.
Nebil GöksuDepartment of Otorhinolaryngology, Faculty of Medicine, Gazi University, Ankara, Turkey.
M Birol UğurDepartment of Otorhinolaryngology, Faculty of Medicine, Gazi University, Ankara, Turkey.
Haluk KülahDepartment of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey. kulah@metu.edu.tr.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Realizations of fully implantable cochlear implants (FICIs) for providing adequate solution to esthetic concerns and frequent battery replacement have lacked of addressing system level criteria as a complete device. Here, we present a full-custom FICI that considers design of both an implantable sensor for wide range sound sensing and a signal conditioning circuit for electrical stimulation of the auditory nerve. The microelectromechanical system (MEMS)-based acoustic sensor utilizes multiple cantilever beam structures to sense and filter the mechanical vibrations on the ossicular chain. The area optimized bilayer design of the piezoelectric sensor met with the volume limitation in the middle ear while achieving high signal-to-noise-ratio. The sensor outputs are processed by a current mode low-power signal conditioning circuit that stimulates the auditory neurons through intracochlear electrodes. The FICI is validated with an in vivo model where the electrical auditory brainstem response (eABR) of the animal was observed while applying sound excitation. The eABR results demonstrate that the system is able to evoke responses in the auditory nerves of a guinea pig for sound range of 45-100 dB SPL within the selected frequency bands.