Evidence map›Paper›PMID 39275742›Full record

ArticleSensors (Basel, Switzerland)2024

High-Accuracy Calibration Method of a Thermal Camera Using Two Reference Blackbodies.

Tomasz Sosnowski, Mariusz Kastek, Krzysztof Sawicki, Andrzej Ligienza, Sławomir Gogler, Bogusław Więcek

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Wound Bed Temperature has Potential to Indicate Infection Status: A Cross-Sectional Study.Wound repair and regeneration : official publication of the Wound Healing Society [and] the European Tissue Repair Society
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Tomasz SosnowskiInstitute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland.ORCID 0000-0003-4082-8366
Mariusz KastekInstitute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland.ORCID 0000-0001-7184-5228
Krzysztof SawickiInstitute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland.ORCID 0000-0002-1368-3854
Andrzej LigienzaInstitute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland.ORCID 0000-0002-5789-8531
Sławomir GoglerInstitute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland.ORCID 0000-0002-3002-6362
Bogusław WięcekInstitute of Electronics, Lodz University of Technology, 90-924 Łódź, Poland.ORCID 0000-0002-5003-1687

Funding

Military University of Technology in Warsaw UGB 464 724/2022National Centre for Research and Development POIR.01.01.01-00-0662/20
6 · The paper itself

Abstract

Body temperature is one of the most important physiological parameters of a human being used to assess his basic vital functions. In medical practice, various types of measuring instruments are used to measure temperature, such as liquid thermometers, electronic thermometers, non-contact ear thermometers, and non-contact forehead thermometers. Such body temperature measurement techniques require the connection of appropriate sensors to a person, and non-contact thermometers operate over short distances and force a specific position of the person during the measurement. As a result, using the above methods, it is practically impossible to perform body temperature measurements of a moving human being. A thermal imaging camera can be used effectively for the purpose of the temperature measurement of moving objects, but the remote measurement of a human body temperature using a thermal imaging camera is affected by many factors that are difficult to control. Accurate remote measurement of human body temperature requires a measurement system that implements a specialized temperature determination algorithm. This article presents a model of a measurement system that facilitates the development of a highly accurate temperature measurement method. For the model, its parameters were determined on the calibration stand. The correct operation of the developed method and the effectiveness of temperature measurement have been confirmed by tests on a test stand using reference radiation sources.

Indexed as

non-contact temperature measurementsignal procesingthermal camera model

Identifiers

PMID39275742
PMCPMC11398128

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.