ArticleJournal of applied clinical medical physics2024
The implementation of low instantaneous dose rate total body irradiation with linear accelerator in small-size treatment rooms.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- The implementation of low instantaneous dose rate total body irradiation with linear accelerator in small-size treatment rooms.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis paper describes the implementation of an instantaneous low-dose-rate total body irradiation (TBI) technique using block-filtered 6 MV X-rays with a linear accelerator (LINAC) to reduce pulmonary toxicity.
methodsIn the absence of dedicated TBI-specific meter-set dose rates in LINAC and sufficient treatment room size, a 2-cm-thick transmission block was used together with a 200-cm source-to-surface distance (SSD) to reduce the instantaneous dose rates of 6 MV x-rays down to 10 cGy/min, thus alteration to the beam properties. A TBI-specific dose calculation model was built with data acquired at the treatment planning system (TPS)-permitted maximum 140-cm SSD and was validated in phantoms at a 180-cm SSD. As for planning strategies, we adopted large anterior-to-posterior/posterior-to-anterior (AP/PA) open fields with multi-leaf collimator shielding for lungs to achieve target coverage, lung protection, and efficient dose delivery. A custom-designed sliding couch (Patent No. ZL202123085880.1) was manufactured to support patients during treatment. Measures to control the quality and safety of TBI treatment include machine interlocks, pretreatment checklists, and in-vivo dose monitoring.
resultsThe instantaneous dose rate of block-filtered 6MV X-ray was reduced to approximately 7.0 cGy/min at 12.5-7.5 cm depth with a 185-200 cm SSD. The dose calculated by TPS differs from the measurements by 0.15%-1.55% in the homogeneous phantom and 1.2%-4.85% in the CIRS thorax phantom. The open-field TBI technique achieved V
conclusionA low instantaneous dose rate TBI technique can be implemented in the clinic.