ArticleOrthopaedic surgery2024
How Does the Stress in the Fixation Device Change during Different Stages of Bone Healing in the Treatment of Fractures? A Finite Element Study of External Fixation for Tibial Fractures.
Article in Orthopaedic surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Staged screw removal in external plate fixation for distal tibial fractures: a finite element study on adaptive biomechanics.Scientific reports · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
backgroundAlthough the specific relationship between the stress changes in the external fixator during tibial fracture treatment and the bone healing process remains unclear, it is believed that stress variations in the external fixator scaffold can, to a certain extent, reflect the progress of tibial healing.
objectiveThis study aims to propose a non-invasive method for assessing the degree of fracture healing by monitoring the changes in stress transmission, the locations of stress-sensitive points, and displacement in the external fixator-tibia system during the healing process of tibial fractures.
methodsIn this study, finite element models of tibial fractures at various healing stages were developed. Physiological conditions, including axial, torsional, and bending loads on the tibia, were simulated to evaluate stress and strain within the external scaffold-tibia system under normal physiological loading conditions.
resultsThe results indicate variations in the stress distribution between the external fixator and the tibia during different stages of healing. In the early phase of fracture healing, the external fixator plays a crucial role as the primary load-bearing unit under all three loading conditions. As the fracture healing progresses, the stress on the tibia gradually increases, concentrating on the medial part of the tibia under axial and torsional loading, and at the upper and lower ends, as well as the central part of the anterior and posterior tibia during bending loading. The stress at the callus gradually increases, while micro-movements decrease. The stress within the external bracket gradually decreases, with a tendency for the connecting rod to transfer stress towards the screws. Throughout the fracture healing process, the location of maximum stress in the external fixator remains unchanged. Under axial and torsional loading, the maximum stress is located at the intersection of the lowest screw and the bone cortex, while under bending loading, it is at the intersection of the second screw and the connecting rod.
conclusionDuring the bone healing process, stress is transferred between the external fixation frame and the bone. As bone healing advances, the stress on the connecting rods and screws of the external fixation frame decreases, and the amplitude of stress changes diminishes. When complete and robust fusion is achieved, stress variations stabilize, and the location of maximum stress on the external fixation frame remains unchanged. The intersections of the lowest screw and the bone cortex, as well as the second screw and the connecting rod, can serve as sensitive points for monitoring the degree of bone healing.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.