Evidence map›Paper›PMID 39210201›Full record

ArticleAnnals of nuclear medicine2024

Validation of advanced hybrid SPECT/CT system using dynamic anthropomorphic cardiac phantom.

Elad Gelbart, Alexander Krakovich, Yigal Sherm, Gilad Rabin, Hagit Ratner, Mickey Scheinowitz, Ronen Goldkorn

Abstract readValidation Study
In one paragraph

Article in Annals of nuclear medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Elad GelbartFaculty of Medical and Health Sciences, Tel-Aviv University, Tel-Aviv, Israel.
Alexander KrakovichDepartment of Biomedical Engineering, Faculty of Engineering, Tel-Aviv University, Tel-Aviv, Israel.
Yigal ShermGE HealthCare Molecular Imaging, Haifa, Israel.
Gilad RabinGE HealthCare Molecular Imaging, Haifa, Israel.
Hagit RatnerGE HealthCare Molecular Imaging, Haifa, Israel.
Mickey ScheinowitzFaculty of Medical and Health Sciences, Tel-Aviv University, Tel-Aviv, Israel.
Ronen GoldkornFaculty of Medical and Health Sciences, Tel-Aviv University, Tel-Aviv, Israel. Ronen.Goldkorn@sheba.health.gov.il.ORCID http://orcid.org/0000-0001-7376-7032

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveMyocardial blood flow (MBF) assessment can provide incremental diagnostic and prognostic information and thus the validation of dynamic SPECT is of high importance. We recently developed a novel cardiac phantom for dynamic SPECT validation and compared its performance against the GE Discovery NM 530c. We now report its use for validation of a new hybrid SPECT/CT System featuring advanced cadmium zinc telluride (CZT) technology in a ring array detector design (StarGuide™, GE HealthCare).

methodsOur recently developed cardiac phantom with injected technetium-99m radiotracer was used to create physiological time activity curves (TACs) for the left ventricular (LV) cavity and the myocardium. The TACs allow the calculation of uptake rate (K1) and MBF. The StarGuide system was used to acquire and process the TACs, and these were compared to the TACs produced by the phantom and its mathematical model. Fifteen (15) experiments with different doses representing various MBF values were conducted, and a standard statistic tool was applied for significance.

resultsThe TACs produced by the StarGuide system had a significant correlation (p < 0.001) with the reference TACs generated by the phantom both for the LV (r = 0.94) and for the myocardium (r = 0.89). The calculated MBF difference between the system and the phantom was 0.14 ± 0.16 ml/min/g and the average relative absolute difference was 13.2 ± 8.1%. A coefficient of variance of ≤ 11% was observed for all MBF subranges. The regional uptake rate values were similar to the global one with a maximum difference of 5%.

conclusionsOur newly developed dynamic cardiac phantom was used for validation of the dynamic hybrid SPECT/CT CZT-based system (StarGuide™, GE). The accuracy and precision of the system for assessing MBF values were high. The new StarGuide system can reliably perform dynamic SPECT acquisitions over a wide range of myocardial perfusion flow rates.

Indexed as

HeartPhantoms, ImagingSingle Photon Emission Computed Tomography Computed TomographyCadmiumCoronary CirculationHumansImage Processing, Computer-AssistedTelluriumZincCadmiumCdZnTeTelluriumZincAnthropomorphic phantomMBFNuclear cardiologySPECTStarGuide

Identifiers

PMID39210201
PMCPMC11489191

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.