Evidence map›Paper›PMID 39206041›Full record

ArticleJournal of mass spectrometry and advances in the clinical lab2024

Comparison between a single- and a multi-point calibration method using LC-MS/MS for measurement of 5-fluorouracil in human plasma.

Mirjana Radovanovic, Jennifer J Schneider, Jennifer H Martin, Ross L G Norris, Peter Galettis

Abstract read
In one paragraph

Article in Journal of mass spectrometry and advances in the clinical lab, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Observational
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Mirjana RadovanovicCentre for Drug Repurposing and Medicines Research, University of Newcastle, Callaghan, NSW, Australia.
Jennifer J SchneiderCentre for Drug Repurposing and Medicines Research, University of Newcastle, Callaghan, NSW, Australia.
Jennifer H MartinCentre for Drug Repurposing and Medicines Research, University of Newcastle, Callaghan, NSW, Australia.
Ross L G NorrisCentre for Drug Repurposing and Medicines Research, University of Newcastle, Callaghan, NSW, Australia.
Peter GalettisCentre for Drug Repurposing and Medicines Research, University of Newcastle, Callaghan, NSW, Australia.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

When quantifying therapeutic drugs using LC-MS/MS instrumentation in clinical laboratories, batch-mode analysis with a calibration curve consisting of 6-10 concentrations for each analyte is the most widely used approach. However, this is an inefficient use of this technology since it increases cost, delays result availability and precludes random instrument access. Various alternative methods to reduce the calibrator use and improve efficiency without compromising analytical quality have been investigated, and a single-point calibration has been reported to be the simplest, least expensive and the quickest approach. This study compares a single and a multi-point calibration method using LC-MS/MS with 5-fluorouracil (5-FU) as a model drug. The method was validated for quantitative analysis of 5-FU over a concentration range of 0.05-50 mg/L. Patients undergoing cancer treatment with intravenous 5-FU had plasma 5-FU concentrations measured, and their dose adjusted in real time based on the calculated area under the time-concentration curve (AUC). Subsequently, a single point calibration method using a concentration at 0.5 mg/L was compared to the multi-point calibration method in terms of accuracy and precision. A Bland-Altman bias plot and a Passing-Bablok regression analysis showed a good agreement between the two methods (mean difference = -1.87 %, slope = 1.002, respectively) when comparing patient plasma 5-FU concentrations. The calibration method did not impact the AUC results nor the decision on 5-FU dose adjustments. Our study demonstrated that a single point calibration method produced analytically and clinically comparable results to those produced by a multi-point method when quantifying 5-FU and is feasible to be used clinically.

Indexed as

5 fluorouracilLC-MSMulti-point calibration methodRandom instrument accessSingle point calibration method

Identifiers

PMID39206041
PMCPMC11350269

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.