Evidence map›Paper›PMID 39143399›Full record

ArticleJournal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology2024

Polyethylene liner dissociation in total hip arthroplasty: a retrospective case-control study on a single implant design.

S Pagano, J F Plate, T Kappenschneider, J Reinhard, M Scharf, G Maderbacher

Abstract read
In one paragraph

Article in Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. For many but not for all: the bikini incision direct anterior approach for total hip arthroplasty. A narrative review.Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology · 2024
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

S PaganoDepartment of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany. stefano.pagano@ukr.de.ORCID http://orcid.org/0009-0006-6721-7564
J F PlateDepartment of Orthopaedic Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
T KappenschneiderDepartment of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany.
J ReinhardDepartment of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany.
M ScharfDepartment of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany.
G MaderbacherDepartment of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundModular acetabular components for total hip arthroplasty (THA) provide intraoperative flexibility; however, polyethylene liner dissociation may occur. This study aimed to examine the incidence and causes of liner dissociation associated with a specific acetabular component design at a single centre. MATERIALS AND

methodsA retrospective analysis of 7027 patients who underwent primary THA was performed to identify isolated liner dislocations. Patient demographics, clinical presentations, surgical and implant details, and both radiographic and computed tomography (CT) findings were analysed. Patients with liner dislocation were matched to a control group via 2:1 propensity score matching, and a logistic regression analysis was employed to identify associated risk factors.

resultsA total of 32 patients (0.45%) experienced liner dislocation at a mean 71.47 ± 60.10 months post surgery. Significant factors contributing to dislocations included the use of a conventional compared with a highly crosslinked polyethylene component (p = 0.049) and screw fixation (p = 0.028). Radiographic and CT analysis highlighted the importance of proper component orientation, revealing that patients experiencing dislocations demonstrated significantly lower acetabular cup anteversion angles (p = 0.001) compared with the control group. Impingement and malposition, identified in 41% and 47% of the cases, respectively, further underscored the multifactorial nature of dislocation risks.

conclusionsWhile the overall rate of polyethylene liner dislocation was low, the findings of this study highlight the importance of appropriate cup placement to decrease the risk of dissociation. It further substantiates the influence of impingement and malposition in liner displacement, with increased mechanical stress exerted on the locking mechanism under adverse conditions and the potential risk increase due to screw placement.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisPolyethyleneProsthesis DesignProsthesis FailureAcetabulumAgedCase-Control StudiesFemaleHumansMaleMiddle AgedPropensity ScoreRetrospective StudiesRisk FactorsTomography, X-Ray ComputedPolyethyleneCase–control studyComponent positioningLiner dislocationModularityPolyethyleneTotal hip arthroplasty

Identifiers

PMID39143399
PMCPMC11324622

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.