Evidence map›Paper›PMID 39137008›Full record

ArticleJournal of applied clinical medical physics2024

Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3.

Jiajian Shen, Xuanfeng Ding, Serdar Charyyev, Xiaoying Liang, Cristina Oancea, Peilong Wang, William G Rule, Wei Liu, Martin Bues, Liyong Lin

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Jiajian ShenDepartment of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.ORCID https://orcid.org/0000-0003-0439-2075
Xuanfeng DingDepartment of Radiation Oncology, Corewell Health Beaumont University Hospital, Royal Oak, Michigan, USA.
Serdar CharyyevDepartment of Radiation Oncology, Stanford University, Palo Alto, California, USA.
Xiaoying LiangDepartment of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.ORCID https://orcid.org/0000-0002-6688-1045
Cristina OanceaADVACAM, Prague, Czech Republic.
Peilong WangDepartment of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.
William G RuleDepartment of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.
Wei LiuDepartment of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.ORCID https://orcid.org/0000-0003-0935-3999
Martin BuesDepartment of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.
Liyong LinDepartment of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking.

methodsA pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty.

resultsThe uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 µs and 1 ms repetition time for the synchrocyclotron accelerator.

conclusionThe AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.

Indexed as

Particle AcceleratorsProton TherapyRadiotherapy DosageRadiotherapy Planning, Computer-AssistedSemiconductorsHumansNeoplasmsProtonsSynchrotronsTime FactorsProtonsdose ratepencil beam scanningproton therapyscanning speedssemiconductor detectorstime structure

Identifiers

PMID39137008
PMCPMC11492390

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked