ArticleCardiovascular and interventional radiology2024
Performance of an Anti-Phase Technology-Powered Microwave Ablation System on Ex Vivo Liver, Lung and Kidney: Analysis of Temperature Trend, Ablation Size and Sphericity.
Article in Cardiovascular and interventional radiology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Roadmap towards personalized approaches and safety considerations in non-ionizing radiation: from dosimetry to therapeutic and diagnostic applications.Physics in medicine and biology · 2026Review
- Efficacy and safety of percutaneous microwave ablation for liver tumors using an antenna with anti-phase technology offering ultraspherical ablation.Radiology and oncology · 2025Article
- Article
Corrections and comments
- Erratum issued
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeInvestigating the performance of the new Dophi™ M150E Microwave Ablation System, in terms of temperature distribution, ablation size and shape, reproducibility. MATERIALS AND
methodsThe Dophi™ M150E Microwave Ablation System was tested on ex vivo liver, lung and kidney, at 6 different settings of time, power and number of MW antennas (single antenna: 50 and 100 W at 5 and 10 min; double antenna: 75 W at 5 and 10 min). The temperature distribution was recorded by Fiber Bragg Grating sensors, placed at different distances from the antennas. The ablation axes were measured and the sphericity index was calculated.
resultsThe standard deviation of ablation axes was < 5 mm, except at the highest energy and time setting for the lung. A maximum temperature rise of ~ 80 °C was measured. The measured ablation axes are overall comparable with the manufacture's values, especially at lower power and with one MW antenna (average maximum difference is 7 mm). The mean sphericity index of 0.95, 0.79 and 0.9 was obtained for the liver, lung and kidney, respectively, with a single antenna. With double antenna setup, the sphericity index was closer to 1 when 75 W for 10 min were used.
conclusionsDophi™ M150E allows good reproducibility of ablation axes for all cases except in the lung at the highest energy level. With one antenna, an almost spherical ablation area for the liver and kidney was obtained. Using double antenna results in more homogeneous temperature distribution within the tissue compared to single antenna.