Performance of an Anti-Phase Technology-Powered Microwave Ablation System on Ex Vivo Liver, Lung and Kidney: Analysis of Temperature Trend, Ablation Size and Sphericity.

Pouya Namakshenas, Tommaso Arcaini, Benedetta Cesare, Alessandro Dorato, Elena Durante, Milena Ricci, Domiziana Santucci, Paola Saccomandi, Eliodoro Faiella

Erratum issuedAbstract read
In one paragraph

Article in Cardiovascular and interventional radiology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Review
  2. Article
  3. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

9 authors.

Pouya NamakshenasDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy.
Tommaso ArcainiDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy.
Benedetta CesareDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy.
Alessandro DoratoDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy.
Elena DuranteDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy.
Milena RicciOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, via Alvaro del Portillo 200, 00128, Rome, Italy.
Domiziana SantucciOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, via Alvaro del Portillo 200, 00128, Rome, Italy.
Paola SaccomandiDepartment of Mechanical Engineering, Politecnico di Milano, via Giuseppe La Masa 1, 20156, Milan, Italy. paola.saccomandi@polimi.it.ORCID http://orcid.org/0000-0003-4236-8033
Eliodoro FaiellaOperative Research Unit of Radiology and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, via Alvaro del Portillo 200, 00128, Rome, Italy.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeInvestigating the performance of the new Dophi™ M150E Microwave Ablation System, in terms of temperature distribution, ablation size and shape, reproducibility. MATERIALS AND

methodsThe Dophi™ M150E Microwave Ablation System was tested on ex vivo liver, lung and kidney, at 6 different settings of time, power and number of MW antennas (single antenna: 50 and 100 W at 5 and 10 min; double antenna: 75 W at 5 and 10 min). The temperature distribution was recorded by Fiber Bragg Grating sensors, placed at different distances from the antennas. The ablation axes were measured and the sphericity index was calculated.

resultsThe standard deviation of ablation axes was < 5 mm, except at the highest energy and time setting for the lung. A maximum temperature rise of ~ 80 °C was measured. The measured ablation axes are overall comparable with the manufacture's values, especially at lower power and with one MW antenna (average maximum difference is 7 mm). The mean sphericity index of 0.95, 0.79 and 0.9 was obtained for the liver, lung and kidney, respectively, with a single antenna. With double antenna setup, the sphericity index was closer to 1 when 75 W for 10 min were used.

conclusionsDophi™ M150E allows good reproducibility of ablation axes for all cases except in the lung at the highest energy level. With one antenna, an almost spherical ablation area for the liver and kidney was obtained. Using double antenna results in more homogeneous temperature distribution within the tissue compared to single antenna.

Indexed as

KidneyLiverLungMicrowavesAblation TechniquesAnimalsCatheter AblationEquipment DesignIn Vitro TechniquesReproducibility of ResultsSwineTemperatureAblation sizeKidneyLiverLungMicrowave ablationSphericityTemperature distribution

Identifiers

PMID39085441
PMCPMC11486822

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.